22
Calibrate
Use the
buttons to move the cursor to the “Option” menu.
From this menu, use the
button to place the cursor on the “Calibr.” option.
Press the Operate button. When the submenu
opens, use the
button to move the cursor to the “Calibrate” option.
Now press the Operate button, and calibration
will begin. The instrument will first call for the
black standard. Place the instrument on the black
standard and press the Operate button.
A beep will indicate that the measure ment has
been completed.
Now place the instrument on the white standard
and press the Operate button.
For gloss calibration, place the instrument on the
high-gloss standard and press the Operate button.
If calibration was performed correctly, the message
Calibration OK
will appear on the display. The instrument returns
to the previous measurement mode.
Use the test with the green standard again to
determine whether the calibration was performed
correctly.