The Fit Anti-Gravity Treadmill will be installed by an AlterG qualified technician after delivery. Please
make sure that you inspect the Anti-Gravity Treadmill upon delivery for any damage that may have
occurred during transportation. Take photographs and report any damage immediately to the shipping
company and AlterG. When you sign for the shipment of your Anti-Gravity Treadmill, you are taking
responsibility for any damage that may occur before installation.
Electrical Requirements
The recommended electrical power connection for the Fit Anti-Gravity Treadmill is a dedicated
20 ampere, 220VAC @ 50/60 Hz circuit with ground. The plug supplied with the Anti-Gravity Treadmill is
designated by the NEMA configuration system as 6-20P
.
The corresponding receptacle for the plug is a
NEMA 6-20R
.
An appropriate plug for the country and facility in question should be wired as follows:
Blue Conductor: Neutral
Brown Conductor: Line
Green/Yellow Conductor: Ground
Grounding Requirements
WARNING: To avoid the risk of electric shock, this equipment must only be connected to
a supply mains with protective earth.
The Fit Anti-Gravity Treadmill must be grounded electrically. If there is an electrical malfunction,
grounding provides a path of least resistance for electric current to reduce the risk of electric shock. The
Anti-Gravity Treadmill power cord includes a grounded plug. This plug must be plugged into an
appropriate receptacle (NEMA 6-20R) that is properly installed and grounded in accordance with the
current National Electrical Code as well as all local codes and ordinances. If you are at all unsure of these
requirements, contact AlterG or a qualified electrician.
Location Requirements
The Fit Anti-Gravity Treadmill measures 84” (213cm)
×
38” (97cm) and weighs almost 550lbs (249kg). It
needs to be placed on a structurally sound surface. If it is used above ground level, it is advisable to place
it near the corner of the room or where the floor will be strongest to ensure maximum support during high-
speed, high-impact use. The surface should be level to ensure minimal flexing of the Anti-Gravity
Treadmill frame. Do not place the Anti-Gravity Treadmill directly on thick carpeting because it may
interfere with air-valves located on the underside of the treadmill. The Anti-Gravity Treadmill needs to be
within 12ft (3.7m) (from the front of the treadmill) of the proper electrical outlet. Check with a qualified
electrician or AlterG if you plan on extending the cord in any way. Make sure you leave at least 24”
(61cm) on either side of the treadmill to allow the bag to expand during inflation. Allow at least 40” (1m)
behind the treadmill to accommodate a user getting in and out safely, and at least 20” (.5m) in front of the
unit to allow access to electronics and power on/off switch.
We recommend an area at least 12ft (3.7m) long by 8ft (2.4m) wide to provide adequate space for
operation and user access. Also, check ceiling height to ensure that users will not hit their heads on the
ceiling while running. The Anti-Gravity Treadmill surface is 6” (15cm) off the floor. An 8ft (2.4m) ceiling
may be too low for taller users.
AlterG FIT Anti-Gravity Treadmill
®
Installations Specifications
•
Maximum height of treadmill: 74” (1.88M)
•
Width of treadmill: 38” (0.97M)
•
Length of treadmill: 84” (2.13M)