SCENARIO 1

 

The patient falls or becomes fatigued to the point of inability to exit the treadmill before the air chamber is inflated with air.

 

Option 1:

  

If the patient can stand, stop the treadmill using the touchscreen controls OR pulling the red emergency stop 

button, then secure the belt with your foot.  Help the patient exit the treadmill.

 

Option 2:

 Prevent the belt from moving by securing it with your foot. Remove the bag rods at the       

 

                   

                                treadmill base.

 

Option 3:

 Prevent the belt from moving by securing it with your foot. Cut the air chamber and help the  

                                  

                                patient exit the treadmill.

 

SCENARIO 2

 

The patient becomes fatigued to the point of inability to exit the treadmill after the air chamber is inflated.

Reduce the body weight percentage to less than 40% and have the patient sit down in place until they re-gain their      

 

strength and are able to exit on their own.

IMPORTANT NOTES: 

1. Do not unlock the cockpit until you read further. 

2. The treadmill belt on the Via series does not lock in place, so secure it by holding your  

    foot against it at the rear of the treadmill.

EMERGENCY EXTRACTION PROTOCOLS

Before using the Anti-Gravity Treadmill, read the entire User Manual, which can be obtained at alterg.com or by contacting our 

Service Department at +1.510.270-5900.  

In the unlikely event that a patient becomes immobilized in the Anti-Gravity Treadmill, please refer to the urgent steps 

below to secure the situation before c1510.270.5900 or referring to alterg.com or the User Manual for more detailed 

instructions and next steps.

EMERGENCY SCENARIOS 

There are two likely scenarios in which a patient may require urgent help with getting out of the Anti-Gravity Treadmill: