Model Q 1074A - OPERATION MANUAL 

 

                                                                                               

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 

4.  How to do Self-Calibration? 

4.1.   Why need to do self-calibration? 

The accuracy of Most DMM is assured under the condition that operating environment is within 
“23±5°C” and the humidity “<75%RH”, otherwise the temperature co-efficiency should be taken into 
consideration. For this DMM, if you perform self-calibration, the accuracy will be kept and the affection 
of environment changes will be eliminated.   
 

4.2.   How to self- calibrate a multimeter? 

1)  Rotate the 

selector switch

 to 

"

Ω/

" position

, make sure the meter is on Ohm mode; 

2)  Press the 

‘Select’ button more than 5 seconds until “CAL” displays on LCD. CAL - means now 

the meter enters into self-calibration status.     

3)  The self-calibration procedure will be finished within 35 seconds. 

 

REMARK: Self-calibration can be done only on a multimeter that has the self-calibration 

function. 

 

 

 

 

 

 

Manual Version 1074-A 

Revised 10/18 Philip Smith 

Screws 

Fuse: 
0.8A/600V FAST

 

Fuse: 
10A/600V FAST

 

Summary of Contents for Q 1074A

Page 1: ...hen servicing the meter use only specified replacement parts Use caution when working above 30 V ac rms 42 V peak or 60 V dc Such voltages pose a shock hazard The finger or any part of your body shall...

Page 2: ...tery voltage is low Auto Power Off 10 minutes after stopping the switch or not pressing any operation button the meter automatically enters to power off mode Safety standards EMC LVD The meter is up t...

Page 3: ...5 10A for 15sec each 15min maximum Overload protect Fast fuse 10A 600V Input voltage drop 0 3V Frequency Respond 50Hz 1KHz Note Not to measure a current in a circuit with voltage more than 600V Capaci...

Page 4: ...ts max then return the min range Press RANGE button more than 2 seconds the meter return to auto range This button do not effect on frequency duty cycle capacitance diode continuity test and temperatu...

Page 5: ...current measuring mode 4 For current measurement from 600mA to 10A connect the red test lead to 10A socket Set selector switch to AC DC 10A position and press SELECT button to select DC or ACA measuri...

Page 6: ...rement 5 Read the result from the LCD panel 2 9 Frequency measurement 1 Connect the black test lead to COM socket and red test leads to the V Hz socket 2 Set the selector switch to Hz position 3 Conne...

Page 7: ...sionally to keep it looking new Do not use harsh chemicals cleaning solvents or strong detergents to clean the multimeter 6 Use only fresh batteries of the required size and type Always remove old or...

Page 8: ...and the affection of environment changes will be eliminated 4 2 How to self calibrate a multimeter 1 Rotate the selector switch to position make sure the meter is on Ohm mode 2 Press the Select butto...

Reviews: