4. How to do Self-Calibration?
4.1. Why need to do self-calibration?
The accuracy of Most DMM is assured under the condition that operating environment is within
“23±5°C” and the humidity “<75%RH”, otherwise the temperature co-efficiency should be taken into
consideration. For this DMM, if you perform self-calibration, the accuracy will be kept and the affection
of environment changes will be eliminated.
4.2. How to self- calibrate a multimeter?
1) Rotate the
selector switch
to
"
Ω/
" position
, make sure the meter is on Ohm mode;
2) Press the
‘Select’ button more than 5 seconds until “CAL” displays on LCD. CAL - means now
the meter enters into self-calibration status.
3) The self-calibration procedure will be finished within 35 seconds.
REMARK: Self-calibration can be done only on a multimeter that has the self-calibration
function.
Manual Version 1074-A
Revised 10/18 Philip Smith