14
CALIBRATION OF TI-PREFORMS BLANK HOLDER
E N
2.4.2 Inserting the Ti-Preforms blank holder
▷
Check the connection mandrel of the blank
holder and clean if required.
▷
Insert the tool holder together with the calibra-
tion sleeve via the connection mandrel to the
stop into the locating bore.
Fig. 2
▷
Tighten fastening screw [1] with the torque
wrench until the torque wrench clicks once.
Fig. 3
Ti-Preforms blank holder
1
Fastening screw for tool holder
2
Tool holder
▷
Carry out a Ti-Preforms blank holder inventory
run.
▷
Insert tools into the tool magazine:
▪
Roto RFID 2.0 Ti (760672)
▪
Roto RFID 1.0 Ti (760673)
The hybrid milling machine carries out an inven-
tory run of the tool magazine.
▷
Start the hybrid milling machine software on the
PC.
▷
Click the
Add job button and switch to the cali-
bration folder by clicking on the
Calibration but-
ton.
▷
Select
M3 Blank Holder Calibration_1.xnc file
and then click on the
Transfer button to trans-
fer data to the hybrid milling machine.
Fig. 4
Transfer
“
M3 Blank Holder
Calibration_1.xnc
”
file
▷
If
Start jobs after transfer was not selected:
Start job manually at the hybrid milling
machine.
The calibration sleeve is now provided with 2
measuring surfaces.
The process takes approx. 20 minutes.