The customer or the user of the system should ensure that it is used in such an 

environment. 

 

NOTE 

These guidelines may not apply in all situations. Electromagnetic propagation is affected 

by absorption and reflection from structures, objects, and people. 

 

Table 2 Electromagnetic Immunity 

Immunity 

Test 

Input/ 

Output 

Port 

IEC 

60601-1-2 

Test 

Level 

Compliance 

Level 

Electromagnetic 

Environment 

Guidance 

Electrostatic 

discharge 

(ESD) 

IEC 61000-4-2 

Enclosure 

port 

±8 kV 

contact 

±15 kV air 

±8 kV contact 

±15 kV air 

The table should be 

wood. If the table is 

covered with 

synthetic material, 

the relative humidity 

should be at least 

30%. 

Power 

Frequency (50/ 

60 Hz) 

magnetic field 

IEC 61000-4-8 

Enclosure 

port 

30 A/m 

30 A/m 

Power frequency 

magnetic fields 

should be at levels 

characteristic of a 

typical location in a 

typical commercial or 

hospital 

environment. 

Radiated RF 

IEC 61000-4-3 

Enclosure 

Port 

10 V/m 

80 MHz to 

2.7 GHz 

10 V/m 

80 MHz to 

2.7 GHz 

For recommended 

separation 

distances, 

see Table 6 

Proximity fields 

from RF 

wireless 

communications 

equipment 

IEC 61000-4-3 

 

Enclosure 

port 

27 V/m 

380 MHz to 

390 

MHz 

 

28 V/m 

430 MHz to 

470 

MHz, 

800 MHz to 

960 

MHz, 

1700 MHz 

to 

Same as 

IEC 60601-1-2 

test level 

Summary of Contents for JHS20W2

Page 1: ...Anhui Huami Healthcare Co Ltd User Manual Version 0 2 Amazfit Health Watch JHS20W2 Anhui Huami Healthcare Co Ltd...

Page 2: ...gned to capture a user s ECG data and transmit the data via Bluetooth to an iOS or Android mobile phone which transmits the ECG data over the network to the Cloud servers The device consists of wristb...

Page 3: ...re the device with another person to avoid cross infection If the device is not to be used for a long time please fully charge the battery and recharge it every 2 months to prevent the battery from ov...

Page 4: ...produce is not allowed to be used together electrosurgical instruments To keep the device away from fire do not heat it otherwise the internal battery may explode 5 Warning MR unsafe This device is N...

Page 5: ...Bluetooth 4 2 Android Operating System versions 4 4 or later iOS 10 0 or later 8 2 Unpack the Midong Health Watch Remove the Midong Health Watch from the box Midong Health Watch 8 3 Charging Charge th...

Page 6: ...tion if necessary 8 6 Connect the Midong Health Watch to mobile phone Turn on the mobile phone Bluetooth and connect the Midong Health Watch following the instructions presented when opening the Amazf...

Page 7: ...just the wristband to the appropriate tightness so that the electrode sensors contact the skin 9 2 3 Rest your right forefinger on the front electrode Press your right forefinger with enough force to...

Page 8: ...l profile open the menu bar by pressing the button on the upper left corner and then click on the portrait area In the personal profile page nickname sex date of birth height and weight can be modifie...

Page 9: ...to change the network environment switch to WiFi 4G 15 Specifications Performance characteristics ECG Channels Single Channel Input Dynamic Range 10 mV Peak to Peak Memory length Practically unlimited...

Page 10: ...it could result in improper operation If such use is necessary this equipment and the other equipment should be observed to verify that they are operating normally In addition to the accessories prov...

Page 11: ...8 kV contact 15 kV air The table should be wood If the table is covered with synthetic material the relative humidity should be at least 30 Power Frequency 50 60 Hz magnetic field IEC 61000 4 8 Enclos...

Page 12: ...y Use only the transducers listed in Table 4 to minimize the chance of performance degradation due to these types of electromagnetic phenomena 17 3Recommended Separation Distance Table 3 provides guid...

Page 13: ...quipment could result 18 Equipment symbols Type BF Applied Part WEEE Properly Dispose of Electronic Waste Consult Instructions for Use User Manual Manufacturer Temperature Limits Operational Relative...

Page 14: ...t generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no gua...

Reviews: