background image

Anhui Huami Healthcare Co.,Ltd. 

Option 4: Try to turn off and on the Bluetooth of the mobile 

phone or restart the mobile phone. 

I have a lot of artifact, 

noise, or interference 

in my recording or no 

ECG recording. 

Option  1:  Ensure  that  your  ECG  Record,  arms,  and  hands 

remain still during recordings. 

Option  2:  Clean  the  electrode  sensors  on  the  device  with  an 

alcohol-based sanitizer. 

Option 3: If your skin is very dry, use a water-based lotion before 

recording. 

Option 4: When recording, relax your arms and hands to reduce 

muscle noise.   

Failure 

in 

data 

uploading 

The data uploading is related to the network environment; you 

can try to change the network environment (switch to WiFi/4G). 

 

15. Specifications   

Performance characteristics 

ECG Channels: Single Channel 

Input Dynamic Range: 10 mV Peak-to-Peak 

Memory length: Practically unlimited 

Recording format: Continuous 

 

Circuitry 

Frequency Resp.: 0.67 Hz to 40 Hz 
CMRR: ≥ 60 dB 
Input impedance: > 10

 

MΩ 

Differential range: ± 5 mV 

A/D Sampling Rate: 250 samples/second 

Resolution: 22 bit 

DC offset correction: 

± 

300

 mV 

 

Input 

Frequency: 2400 - 2483.5 MHz 

Modulation: GFSK 

Frequency: 2400~2483.5MHz 

MAX Transmitted power:  dBm 

 

Power requirement 

AC Input: 100V - 250V, 50Hz/60Hz to 5VDC Adapter with 5V USB Adapter Cable 

Battery Type: DC 3.85V, 190 mAh, Lithium battery 

Charging Time: 2.5 hours from flat 

Operating Time: 24 hours on a full charge 

 

Physical characteristics 

Midong Health Watch: 42 × 35 × 12.5 mm; 25 grams 

Wristband length: 118~195 mm 

0

Summary of Contents for JHS20W2

Page 1: ...Anhui Huami Healthcare Co Ltd User Manual Version 0 2 Amazfit Health Watch JHS20W2 Anhui Huami Healthcare Co Ltd...

Page 2: ...gned to capture a user s ECG data and transmit the data via Bluetooth to an iOS or Android mobile phone which transmits the ECG data over the network to the Cloud servers The device consists of wristb...

Page 3: ...re the device with another person to avoid cross infection If the device is not to be used for a long time please fully charge the battery and recharge it every 2 months to prevent the battery from ov...

Page 4: ...produce is not allowed to be used together electrosurgical instruments To keep the device away from fire do not heat it otherwise the internal battery may explode 5 Warning MR unsafe This device is N...

Page 5: ...Bluetooth 4 2 Android Operating System versions 4 4 or later iOS 10 0 or later 8 2 Unpack the Midong Health Watch Remove the Midong Health Watch from the box Midong Health Watch 8 3 Charging Charge th...

Page 6: ...tion if necessary 8 6 Connect the Midong Health Watch to mobile phone Turn on the mobile phone Bluetooth and connect the Midong Health Watch following the instructions presented when opening the Amazf...

Page 7: ...just the wristband to the appropriate tightness so that the electrode sensors contact the skin 9 2 3 Rest your right forefinger on the front electrode Press your right forefinger with enough force to...

Page 8: ...l profile open the menu bar by pressing the button on the upper left corner and then click on the portrait area In the personal profile page nickname sex date of birth height and weight can be modifie...

Page 9: ...to change the network environment switch to WiFi 4G 15 Specifications Performance characteristics ECG Channels Single Channel Input Dynamic Range 10 mV Peak to Peak Memory length Practically unlimited...

Page 10: ...it could result in improper operation If such use is necessary this equipment and the other equipment should be observed to verify that they are operating normally In addition to the accessories prov...

Page 11: ...8 kV contact 15 kV air The table should be wood If the table is covered with synthetic material the relative humidity should be at least 30 Power Frequency 50 60 Hz magnetic field IEC 61000 4 8 Enclos...

Page 12: ...y Use only the transducers listed in Table 4 to minimize the chance of performance degradation due to these types of electromagnetic phenomena 17 3Recommended Separation Distance Table 3 provides guid...

Page 13: ...quipment could result 18 Equipment symbols Type BF Applied Part WEEE Properly Dispose of Electronic Waste Consult Instructions for Use User Manual Manufacturer Temperature Limits Operational Relative...

Page 14: ...t generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no gua...

Reviews: