22
-
●
Lid lock temperature
The temperature that prevents the load from being taken out for safety can be set for each mode
and each program number. The initial setting of the lid lock is 80
℃
for the liquid, agar, and
dissolution modes, while it is 97
℃
for the solid mode.
The set value cannot be changed during operation.
●
The center temperature of liquid will take much time to drop. Take the delay in temperature drop
into sufficient consideration when setting the lid lock temperature in order to avoid an accidental
scald.
①
Press the "FUNC." switch three times in the standby state.
・
"F3" will appear on the digital display, and the lid lock set temperature display will blink, entering
the changeable state.
②
Press the set value increase/decrease switch (
▲
▼
) to change the value.
・
Every time the switch is pressed, the temperature increases (decreases) in 1
℃
increments
within the following range.
HG-50 : within the range of 60
℃
to 97
℃
HG-80 : within the range of 60
℃
to 95
℃
for liquid, agar and dissolution modes
within the range of 60
℃
to 97
℃
for solid mode.
・
While the switch is kept pressed, the set value increases(decreases) in 10
℃
increments and
if the value exceeds the upper (lower) limit, it will return to the lower (upper) limit.
③
If the switch is left untouched for 5 seconds, the display will return to the standby state,
and completing the setting .
CAUTION
Summary of Contents for Hirayama HG-50
Page 8: ... 7 HG 80 External View ...