8. Starting an Operation
①
Press the "START" switch.
・
The lid will close automatically, the
display will light up, and the stage
shown below will begin automatically according to the selected cycle.
Refer to "13. Operation of each stage" on page 27 for the details of the
operation of each stage.
・
When operation will not begin even if the "START" switch is pressed, check to
see if the drain bottle is full of water. If it is full, refer to "1. Draining the Drain
Bottle" on page 31. Operation will not begin either when the lid has not been
opened or closed completely. Press the "LID OPEN" switch, and press the
“START” switch again after the lid has opened completely.
●
When closing the lid, be careful so that a hand will not be caught in the lid.
●
Do not touch the lid or lid cover when opening/closing the lid.
●
Check that there is no foreign material on the mating face of the gasket before starting the
operation. Such a foreign material will cause steam leakage.
●
Do not apply force to the lid that is open. Such force may damage the product.
Liquid mode
Agar mode
Solid mode
Dissolution mode
Lid close
▼
▼
▼
▼
Programmable auto-start operation
▼
▼
▼
▼
Air release
▼
▼
▼
▼
▼
Heating
Heating
▼
▼
▼
▼
Sterilization
Dissolution
▼
▼
▼
▼
Exhaust/cooling
▼
▼
▼
▼
▼
Cooling
▼
▼
▼
▼
▼
Keeping-warm
▼
Keeping-warm
▼
▼
▼
▼
Completion
●
When the lid is closed while the temperature inside the chamber is still high, the air in the
chamber leaking through the gasket will make a hissing sound, which is not failure.
Continue operation.
●
To confirm the set temperature and time during operation, press the "SET" switch.