29
-
■
Exhaust/cooling stage
……
Liquid mode, agar mode, and solid mode
・
The illuminated exhaust section of the stage display changes to blinking, and the pressure inside
the chamber appears on the digital display.
・
The exhaust valve opens up to the preset exhaust level, and the fan begins to run when the fan
cooling is set in the ON position.
・
The liquid mode and agar mode permit the change of exhaust level and fan cooling ON/OFF even
during the exhaust/cooling stage.
●
Rapid exhaust after sterilization of liquid may cause the liquid to spill. Set a lower exhaust level
for fine exhaust, or set the level 0 (natural cooling).
・
When the pressure in the chamber drops below 0.01 MPa and the temperature in the chamber
below 99.9
℃
, the operation proceeds to the next cooling stage.
■
Cooling stage
……
Common to all modes
・
The illuminated cooling section of the stage display changes to blinking, and the exhaust valve
opens fully.
・
The fan runs when the fan cooling is set in the ON position. The fan cooling ON/OFF can be
changed in any stage except the solid mode.
・
The liquid mode and solid mode shift to the next stage when the temperature inside the chamber
drops below the lid lock temperature. The fan continues to run until the temperature drops to the
level 10
℃
lower than the lid lock temperature.
・
The agar mode and dissolution mode permit lid opening/closing when the temperature inside the
chamber drops below the lid lock temperature. When the temperature inside the chamber drops
lower than the warming set temperature, the fan stops, shifting to the next warming stage.
■
Warming stage
……
Mode 1 and 4
・
The illuminated warming section of the stage display changes to blinking.
・
The electronic alarm (beep beep beep) notifies the warming start.
・
When 24 hours (fixed) passes, the operation shifts to the next completion process.
●
After warming time is over (24 hr), the chamber will not be heated. The temperature inside the
chamber will drop to the room temperature, and the agar media, etc. will solidify.
●
Refer to "9. Taking Out the Load" on page 26 when taking out the loaded objects during warming
stage.
△
NOTE
△
NOTE
Summary of Contents for Hirayama HG-50