34
-
Chapter 5. Specifications
Model
HG-50
HG-80
Outer dimensions
455W
×
890H
×
691D mm
455W
×
1030H
×
691D mm
Chamber size
364 Dia.
×
482 D mm
(Effective volume: 50.2 liter)
364 Dia.
×
730 D mm
(Effective volume: 76 liter)
Rated power
AC220 V
±
10%;
Φ
1; 50/60 Hz; 14 A or more
AC230 V
±
10%;
Φ
1; 50/60 Hz; 14 A or more
AC240 V
±
10%;
Φ
1; 50/60 Hz; 13 A or more
AC220 V
±
10%;
Φ
1; 50/60 Hz; 17.3 A or more
AC230 V
±
10%;
Φ
1; 50/60 Hz; 16.6 A or more
AC240 V
±
10%;
Φ
1; 50/60 Hz; 15.9 A or more
Temp./humidity conditions
5-35
℃
, 10-85
%
RH (Dew concentration not allowed.)
Power consumption
3.0kW
3.8kW
Weight
62
㎏
73kg
Type of pressure vessel
Small pressure vessel
Material of chamber
Stainless steel (SUS304)
Sterilization temperature
setting range
105-135
℃
variable
Sterilization timer
1-300 min. (5 hr)/remaining time display method
Dissolution temperature
setting range
60-100
℃
variable
Dissolution timer
1-60 min. (1 hr)/remaining time display method
Warming
temperature
setting range
40-60
℃
variable
Exhaust level
4 steps variable
Fan cooling
ON
/
OFF
Air release time
5 - 10 min. variable
8-16 min. variable
Lid lock temperature
60-97
℃
variable
Liquid, Agar, Dissolution modes: 60-95
℃
variable, Solid mode:60-97
℃
variable
Programmable auto-start
1-99 hr later/ start time setting method
Max. operating pressure
0.26MPa
Thermometer
Digital display: 5.0-137.9
℃
Pressure gauge
Digital display: 0-0.3 MPa/ Analog display: 0-0.4 MPa
Safety alarm unit
Pressure safety valve, Earth leakage and Over current breaker, Boil-dry protection device,
Error display (Boil-dry, Temperature sensor disconnection, Overheating, Overcooling,
Abnormal pressure, Abnormal state of the lid, exhaust valve or heater)
Supplied Accessories
Drain Bottle (1 pc)
Bottom Plate (1 pc)
Operation Manual (1 copy)
Caster Stopper (2 pcs)
Drain Bottle (1 pc)
Bottom Plate (1 pc)
Operation Manual (1 copy)
Caster Stopper (2 pcs)
For HG-50 with 2kW heater as option, the specifications are as follows.
Other specifications are the same as the HG-50 standard model.
Rated power supply
AC110 V
±
10%; single-phase; 50/60 Hz; 19 A or more
AC120 V
±
10%; single-phase; 50/60 Hz; 17 A or more
AC220 V
±
10%; single-phase; 50/60 Hz; 10 A or more
AC230 V
±
10%; single-phase; 50/60 Hz; 9 A or more
AC240 V
±
10%; single-phase; 50/60 Hz; 9 A or more
Air release time
3 - 9 min. variable
Summary of Contents for Hirayama HG-50