Last Alarm/Event Packet
field in left pane (shows date and time)
•
Activity, Alerts
(if the alert is enabled), and/or
Incoming Packets
tabs
The Triton relay (
Triton Relay
) is normally in an Open state and only changes state when taking a measurement.
If an over-current condition occurs, preventing the Triton relay from closing during the measurement, the relay
will go into an alarm state. A packet with data log is sent to the Bullhorn Web account, and the
Triton AC
Coupon Current Density
reading be invalid. The RM4210 will continue to retry taking a measurement.
Once the
overcurrent condition has been corrected, the next scheduled measurement will succeed, and the relay’s state
will return to normal.
The Triton relay does not need to be reset.
The packets that are sent to your Bullhorn Web account can be viewed on the
Unit Detail
page from any of these
sources:
•
Last Packet Type
field in left pane
•
Last Alarm/Event Packet
field in left pane (shows date and time)
•
Activity, Alerts
(if the alert is enabled), and/or
Incoming Packets
tabs
Complete the following steps to reset the test coupon relay (
Test Cpn Relay
) after any over-current issues have
been resolved:
N
OTE
:
The reset over-current function only applies to the test coupon relay; the Triton relay is not affected
when you reset over-current.
1
Click the
Reset Over-Current
button or click
Action > Reset Over-Current
from the main menu.
2
Click
OK
when the
Reset Over-Current Complete
dialog box displays.
Figure 3-25. Reset Over-Current Complete
The indicator bubble next to
Test Cpn Relay
will return to gray, and the state will be
CLOSED
.
N
OTE
:
If the over-current issue was not resolved before resetting the relay, the relay will return to an Open state
and remain in alarm. The over-current condition must be resolved before the relay can be reset to Closed.