Rev. 5
93
Operation
External Rampdown : External Rampdown while in Persistent Mode
1. The Model 430 Programmer first ramps the power supply to the
magnet current.
2. Once the power supply is at the magnet current, the
FIELD AT
TARGET
LED will light and the unit will momentarily “hold”:
3. Following a short “hold”, the persistent switch will be heated to
place the power supply in control of magnet current:
4. After heating (turning off) the persistent switch, the
MAGNET IN
PERSISTENT MODE
LED extinguishes, and rampdown begins.
The
FIELD AT TARGET
LED extinguishes.
5. After rampdown, the following will display:
Note
The external rampdown signal MUST be cleared before pressing
ENTER
will yield a response. Further operator control is inhibited
until the external rampdown signal is cleared.
6. Once the external signal has been cleared,
ENTER
can be pressed.
User control will be re-established and the operator can continue
+3.92 A
Mode: Ramping
+0.17 Vs
PSwitch Heater: OFF
+50.00 A -
Mode: Holding
+2.11 Vs
PSwitch Heater: OFF
+50.00 A
Mode: Heating Switch (4)
+2.11 Vs
PSwitch Heater: ON
+48.85 A
Mode: Zeroing Current
+1.89 Vs
PSwitch Heater: ON
+0.00 A -
Ext. Rampdown completed
-0.00 Vs
Press ENTER to continue
Summary of Contents for 05100PS-430-601
Page 2: ......
Page 10: ...x Rev 5 List of Figures ...
Page 12: ...xii Rev 5 List of Tables ...
Page 18: ...xviii Rev 5 Foreword Safety Summary ...
Page 30: ...12 Rev 5 Introduction Operating Characteristics ...
Page 42: ...24 Rev 3 Installation Power Up Procedure ...
Page 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Page 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Page 168: ...150 Rev 5 Service Return Authorization ...
Page 226: ...208 Rev 5 Index ...