76
Rev. 5
Operation
Example Setup
Note
This item is only available in the Net Setup submenu if IP Address
Assignment is Static. If IP Address Assignment is DHCP, the
gateway IP address is assigned by the network DHCP server and
cannot be assigned by the user.
3.11 Example Setup
As a precursor to operating a superconducting magnet with the Model 430
Programmer and power supply, all of the setup items should be reviewed
and set if necessary with appropriate values for the connected
superconducting magnet.
Figure 3-13 (on the next page) shows an example magnet specifications
sheet. Several parameters needed to operate the magnet are specified.
These values should be entered into the appropriate setup menu of the
Model 430 Programmer. For the purposes of this example, the AMI Model
4Q06125PS power supply will be assumed, since rated current for the
example magnet is 85.6 A.