Rev. 5
85
Operation
Persistent Switch Control : Exiting Persistent Mode
Note
If desired, press the
ESC
key return the Model 430 Programmer to
the default display.
Note
Refer to section 3.6.2 on page 35 for the procedure to display the
magnet current that was established when the persistent switch was
cooled.
6. If desired, the power supply system can be de-energized. Turn the
power supply off first followed in a few seconds by the Model 430
Programmer.
Note
The Model 430 Programmer will store the state of the magnet in
memory and assist the user in exiting the persistent mode when the
Model 430 Programmer is next turned on (discussed in the section to
follow).
3.13.3 Procedure for Exiting Persistent Mode
To exit the persistent mode of magnet operation, the user should perform
the following steps:
1. If the Model 430 Programmer
has not been
powered off since the
magnet was placed in persistent mode,
proceed to step 3
.
2. If the Model 430 Programmer
has been
powered off since the
magnet was placed in persistent mode, complete the following
steps a. and b. before proceeding to step 3.
a. Energize the Model 430 and wait for the prompt on the Model
430 display and then energize the power supply.
After the power supply has been on for a few seconds, press
ENTER
to clear the Model 430 screen prompt.
+0.00 A -
Mode: Zero Current
+0.00 Vm
PSwitch Heater: OFF
+0.00 A
P
Turn on power supply
+0.00 Vs
Press ENTER to continue
Summary of Contents for 05100PS-430-601
Page 2: ......
Page 10: ...x Rev 5 List of Figures ...
Page 12: ...xii Rev 5 List of Tables ...
Page 18: ...xviii Rev 5 Foreword Safety Summary ...
Page 30: ...12 Rev 5 Introduction Operating Characteristics ...
Page 42: ...24 Rev 3 Installation Power Up Procedure ...
Page 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Page 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Page 168: ...150 Rev 5 Service Return Authorization ...
Page 226: ...208 Rev 5 Index ...