180
R
EV
9
A
MERICAN
M
AGNETICS
, I
NC
.
R
EMOTE
I
NTERFACE
R
EFERENCE
: E
RROR
M
ESSAGES
•
-109,”N/A in present mode”
The command or query is not applicable to the present operational
mode. For example, in Short Sample Mode the commands or queries
associated with operation of the Persistent Switch do not apply.
•
-110,”Zero inductance”
The command or query is not available due to a zero inductance value in
the Load Setup submenu.
Q
UERY
E
RRORS
•
-201,”Unrecognized query”
The query string sent (identified as a query by a
?
) was not identified as
valid. Check the query string for invalid characters or separators, syntax
errors, or for errors in the mnemonics. Spaces are not allowed before or
after colon separators.
•
-202,”Undefined coil const”
The user attempted to invoke a query with units of field without first
setting a value for the coil constant. The coil constant must be a non-
zero, positive value.
•
-203,”Query interrupted”
A new query was processed before the return string of a previous query
had been completely transmitted to the host. The new query clears the
remaining data and replaces it with the new return string.
•
-204,”No recorded events”
There are no events in the request for output of the quench or rampdown
standard or backup event files.
E
XECUTION
E
RRORS
•
-301,”Heating(Cooling) switch”
The user attempted to initiate a disallowed function during the persistent
switch heating or cooling transition period. Ramping functions, for
example, are disallowed during the transition period.
•
-302,”Quench condition”
The user attempted to change the ramping state while a quench
condition was active. A quench condition must be cleared via the
RESET
QUENCH
SHIFT-key or by remote command before the ramping state
can be modified.
Summary of Contents for 430
Page 2: ......
Page 16: ...XVI REV 9 AMERICAN MAGNETICS INC FOREWORD SAFETY SUMMARY ...
Page 28: ...12 REV 9 AMERICAN MAGNETICS INC INTRODUCTION OPERATING CHARACTERISTICS ...
Page 64: ...48 REV 9 AMERICAN MAGNETICS INC INSTALLATION POWER UP AND TEST PROCEDURE ...
Page 208: ...192 REV 9 AMERICAN MAGNETICS INC SERVICE RETURN AUTHORIZATION ...
Page 248: ...232 REV 9 AMERICAN MAGNETICS INC APPENDIX SHORT SAMPLE MODE ...