R
EV
9
197
A
MERICAN
M
AGNETICS
, I
NC
.
A
PPENDIX
: C
URRENT
T
RANSDUCER
C
ONNECTOR
C
URRENT
T
RANSDUCER
C
ONNECTOR
CAUTION
Operating the system without the connection between the
Model 430 and the current transducer (CT) can result in loss
of control, and may damage the CT.
The current transducer connector, which is only present for the High-
Stability Option or systems with a maximum current greater than 250 A,
provides pins for connection of the external current transducer (CT) to
the Model 430 Programmer. Power to the CT is also provided via this
connector. The connector is a 9-pin D-sub female connector with the
shell lugs connected to the Model 430 Programmer chassis ground.
Current Transducer Connector Pin Definitions
Pin
Function
1
Sense Out
−
2
not used
3
not used
4
Power Input: 0 VDC (common)
5
Power Input:
−
15 VDC
6
Sense Out +
7
not used
8
not used
9
Power Input: +15 VDC