Use of the
STAB.
(SHIFT+9) key provides a shortcut to the menu for
defining the Model 430 stability setting. The stability setting is specified
in percent and controls the transient response and stability of the
system. The valid input range is from 0.0 to 100.0%. The default value is
0.0% unless preset by AMI to match a specific superconducting magnet.
See page 79 for details of how to determine the stability setting to use.
V
S
<> V
M
(S
HIFT
+
0)
The
Vs <> Vm
(SHIFT+0) key is used to toggle the voltage display
between display of the voltage across the magnet (Vm) and the power
supply output voltage (Vs). See page 52 for details.
V
OLT
M
ETER
(S
HIFT
+
.)
The
VOLT METER
(SHIFT+.) key is used to toggle the main display
between display of a voltmeter indicating magnet voltage (Vm) or supply
voltage (Vs), and display of ramp mode and persistent switch heater
state. See page 53 for details.
F
INE
A
DJUST
(
S
HIFT
+/-
)
The
FINE ADJUST
(SHIFT +/-) key is used to enable the use of the
front panel fine adjust knob to adjust numeric values. See page 55 for
details.
S
HIFT
+ P
ERSIST
.
S
WITCH
C
ONTROL
This keystroke enables persistent switch control
without
some automatic
sequencing features. Refer to page 58.
Summary of Contents for 430
Page 2: ......
Page 16: ...XVI REV 9 AMERICAN MAGNETICS INC FOREWORD SAFETY SUMMARY ...
Page 28: ...12 REV 9 AMERICAN MAGNETICS INC INTRODUCTION OPERATING CHARACTERISTICS ...
Page 64: ...48 REV 9 AMERICAN MAGNETICS INC INSTALLATION POWER UP AND TEST PROCEDURE ...
Page 208: ...192 REV 9 AMERICAN MAGNETICS INC SERVICE RETURN AUTHORIZATION ...
Page 248: ...232 REV 9 AMERICAN MAGNETICS INC APPENDIX SHORT SAMPLE MODE ...