events reaches 100, the standard quench events file becomes the
backup file, and a new (empty) standard quench events file is created.
The standard quench events file contains data from the most recent
quench events, and the backup file (if it exists) contains data from the
100 quench events preceding the oldest record in the standard quench
events file.
The output of this query is terminated with two contiguous pairs of
<CR><LF> characters (i.e. double terminators).
NOTE
The
QUenchFile?
and
QUenchBackup?
queries can return
thousands of ASCII characters. Ensure your input communicaton
buffers are setup to handle a large amount of text.
R
AMPDOWN
S
TATE
C
OMMANDS
AND
Q
UERIES
The external rampdown function also has ramp segmenting capability—
the function operates in a manner similar to the normal ramp as
described on page 63, but the parameters
can only be edited via the
remote interface
.
•
CONFigure:RAMPDown:RATE:SEGments <
# segments
>
Sets the number of external rampdown segments.
•
RAMPDown:RATE:SEGments?
Returns the number of external rampdown segments.
•
CONFigure:RAMPDown:RATE:CURRent <
segment
>,<
rate (A/s,
A/min)
>, <
upper bound (A)
>
Sets the external rampdown rate for the specified segment (values of 1
through the defined number of rampdown segments are valid) in units of
A/sec or A/min (per the selected rampdown rate units), and defines the
current upper bound for that segment in amperes.
•
RAMPDown:RATE:CURRent: <
segment
>?
Returns the external rampdown rate setting for the specified segment
(values of 1 through the defined number of rampdown segments are
valid) in units of A/sec or A/min (per the selected rampdown rate units)
and the current upper bound for that range in amperes. The two return
values are separated by a comma. For example:
Sent:
RAMPDown:RATE:CURRENT:1?
Returned:
0.1000,50.0000
Summary of Contents for 430
Page 2: ......
Page 16: ...XVI REV 9 AMERICAN MAGNETICS INC FOREWORD SAFETY SUMMARY ...
Page 28: ...12 REV 9 AMERICAN MAGNETICS INC INTRODUCTION OPERATING CHARACTERISTICS ...
Page 64: ...48 REV 9 AMERICAN MAGNETICS INC INSTALLATION POWER UP AND TEST PROCEDURE ...
Page 208: ...192 REV 9 AMERICAN MAGNETICS INC SERVICE RETURN AUTHORIZATION ...
Page 248: ...232 REV 9 AMERICAN MAGNETICS INC APPENDIX SHORT SAMPLE MODE ...