background image

D. CALIBRATION (continued)

6. Direct Calibration of the K Factor

This procedure is useful if the meter is over-dispensing or under-

dispensing fluid by a known percentage. Using this method, the 

correction of the USER K FACTOR must be calculated by the 

operator using the following formula:

 New Cal. Factor    =    Old Cal Factor  *  

(       )

100 - E%

100

Example:

Error percentage found

 (

E%)

 

- 0.3 %

CURRENT calibration factor

  1.000

New USER K FACTOR

  

1.000 * [(100 – ( - 0.3))/100] 

= 1.000 * [(100 + 0.9)/100] 

= 1.003

If  the  TIM-615  is  under-dispensing  (negative  error),  the  new 

calibration factor must be higher than the old one as shown in the 

example. The opposite applies if the TIM-615 is over-dispensing 

(positive error).

ACTION

DISPLAY 

CONFIGURATION

1

Make sure the meter is in standby mode

.

If you are unsure, press the REST button and wait 10 

seconds.

2

Hold  down  the  CAL  button  until  the  display  reads 

“CAL”.

The  TIM-615  enters  calibration  mode,  shows  <<CAL>> 

and  displays  the  calibration  factor  in  use.  The  words 

“FACT”  and  “USER”  indicate  which  of  the  two  factors 

(factory or user) is currently in use.

3

Hold down the RESET button until the display reads 

“FIELD”.

The TIM-615 shows “CAL” and the batch register set to 

zero. The TIM-615 is ready to perform in-field dispensing 

calibration.

4

Hold down the RESET button until the display reads 

“DIRECT”.

In  the  bottom  left  part  of  the  display  an  arrow  appears     

(pointing upwards or downwards), that shows the direction 

(increase or decrease) that the value on the batch display 

will be changed when performing step 6.

5

Changing the Arrows direction in the display

Quickly pressing the RESET button changes the arrows 

direction in the display. This operation can be repeated to 

alternate the direction of the arrow. 

6

Adjusting the Batch Display Value (

The batch display 

value changes in the direction indicated by the arrow)

1) The amount on the display changes one unit each 

   time the CAL button is pressed quickly. 

2) The amount on the display changes continuously 

if  the  CAL  button  is  kept  pressed.  The  speed

 

   of  change  increases  the  longer  the  button  is 

held.

If you accidentally program the wrong value, repeat the 

operation starting from step number 5.

7

Saving the New Calibration Factor

Before performing this operation, double-check to make 

sure the display value on the meter is equal to the value 

you obtained using the calibration formula.

Press the RESET button for at least one second. The TIM-

615 calculates the new USER K FACTOR. This calculation 

could require a few seconds to compute. 

At the end of the calculation, the new USER K FACTOR is 

shown for a few seconds. 

8

Finishing

The meter will re-start and enter into the standby mode. 

The meter is now programmed with  the new calibration 

factor and is ready to use.

Note:  The  calibrated  USER  K  FACTOR  is  now  the 

factor used in the TIM-615. This will be the factor the 

meter  will  use  even  after  battery  failure  and  battery 

replacement.

E. UNIT OF MEASURE SETUP

The user can select the batch measurement unit. The choices are: 

Quarts (

Qts

), Pints (

Pts

), Liters (

Lit

), and Gallons (

Gal

). The total 

register unit of measurement is automatically determined based on 

what is chosen for the batch measurement unit (see chart below).

Combination no.

Unit of Measurement

Partial Register

Unit of Measurement

Totals Register

1

Liters (L)

Liters (L)

2

Gallons (Gal)

Gallons  (Gal)

3

Quarts (Qts)

Gallons (Gal)

4

Pints (Pts)

Gallons (Gal)

E. UNIT OF MEASURE SETUP (continued)

To choose between the 4 available combinations:

Wait  for  the  TIM-615  to  go  to 

Standby.

Press and hold the CAL and RESET 

buttons   together. Keep these pressed 

until the word “UNIT” appears on the 

screen. You will also see the current 

units of measure (in this example Qts 

& Gal).

Each  quick  press  of  the  RESET 

button changes the units of measure. 

The  chart  below  shows  the  order  of 

scrolling.

Press the CAL key for a few seconds 

and  the  new  unit  of  measurement 

will be stored. The meter will re-start 

and    enter  into  the  standby  mode. 

The meter is now programmed with 

the  new  unit(s)  of  measure  and  is 

ready to use.

F. METER MAINTENANCE

The TIM-615 was designed to require minimal maintenance. The 

only maintenance required is battery replacement and cleaning the 

measuring chamber. 

1) Battery Replacement

The  TIM-615  is  supplied  with  two  SIZE AAA,  1.5  Volt Alkaline 

batteries. The meter features two low-battery alarm warnings:

When  the  battery  charge  falls   

below the first level, the fixed battery 

symbol  appears  on  the  LCD.  In 

this condition, the meter continues 

to  operate  correctly,  but  the  fixed 

battery icon warns the user that it is 

time to change the batteries. 

If  you  continue  to  use  the  meter 

without  changing  the  batteries 

the  second  battery  alarm  level  will 

eventually  be  reached,  preventing 

meter  operation.  In  this  condition 

the  battery  icon  will  flash  and  will 

be the only item visible on the LCD. 

You must change the batteries once 

the  meter  has  reached  this  mode. 

All  dispensing  data  from  this  point 

forward  will  not  be  displayed  or 

stored in the meter.

To change the batteries, refer to the spare parts list 

in the next column while completing the following 

steps:

Press reset to update all the totals.

Remove the 4 screws on the cover (item 5).

Remove the cover (item 8).

Replace the two dead batteries (item 9).

Put the cover on and re-tighten the 4 cover screws, 

making sure that it has been positioned correctly.

The meter will switch on automatically and enter 

into the standby mode. It is now ready to resume 

normal operation.

After changing the batteries, the meter will display 

the  same  reset  total,  the  same  non-ressettable 

total and the same batch total as it did before the 

batteries were changed. The meter will also use the 

same calibration factor as it did before the power 

outage/battery replacement.

2) Cleaning

The measuring chamber of the TIM-615 can be cleaned without 

removing the unit from the dispensing nozzle or the line on which 

it is installed.

To clean debris from the measuring chamber, refer to the spare 

parts list on the right while completing the following steps:

Loosen  the  six  retention  screws  (item  7)  on  the 

body (item 1).

Remove the cover (item 2) and the seal (item 10)

Remove the two oval gears (item 13).

Clean where necessary. For this operation, use a 

brush or pointed object such as a small screwdriver. 

Be careful not to damage the body or the gears. 

To  reassemble  the  unit,  carry  out  these  steps  in 

the reverse order. Please pay close attention to the 

note below before re-assembling the meter!

G. TROUBLESHOOTING GUIDE

Problem

Possible cause

Remedial Action

Dull LCD

Low Battery

Replace the batteries 

(see section F.)

Inaccurate 

Dispensing

Wrong K FACTOR

Re-calibrate 

(see section D.)

Reduced or Zero 

Flow Rate

Gears blocked  

Clean the measurement 

chamber (see section F.)

The meter does 

not count, but 

the flow rate is 

correct

Incorrect installation of 

gears after cleaning

Repeat the reassembly 

procedure 

(see section F.)

Possible circuit board 

problem

Contact your local 

American Lubrication 

Equipment dealer

H. PARTS BREAKDOWN & SPECIFICATIONS

Item

Part Number

Description

Qty

1

TIM-615-1

Fluid Chamber Body

1

2

TIM-615-2

Isolation Plate

1

3

TIM-615-3

Battery Support Tray

1

4

TIM-615-4

Circuit Board

1

5

TIM-615-5

Cover Screw

4

6

TIM-615-6

Circuit Board Screw

4

7

TIM-615-7

Fluid Body Screw

6

8

TIM-615-8

Protective Cover

1

9

TIM-615-9

AAA Battery

2

10

TIM-615-10

Gasket

1

11

TIM-615-11

Cover label

1

12

TIM-615-12

Adhesive Strip

2

13

TIM-615-13

Magnetic Gear

1

14

TIM-615-14

Non-Magnetic Gear

1

15

TIM-615-15

Gear Post

2

16

TIM-615-HEAD

Replacement Meter 

Head includes items 2, 

3, 4, 5, 6, 8 , 9, 11 & 12

1

H. PARTS BREAKDOWN & SPECIFICATIONS (continued)

Measurement 

system

Oval Gear

Resolution

(nominal)

0.005  (Quarts/Pulse)

Flow Rate

(Range)

1-26  (Quarts/Minute)

Operating pressure

(Max)

1000 (PSI)

Bursting pressure

(Min)

2000 (PSI)

Storage temperature

(Range)

-5 to 158  (°F)

Storage humidity

(Max)

95  (% RH)

Operating 

temperature

(Max)

140  (°F)

Flow resistance

(at 16 quarts/min with 

SAE10W motor oil at 68°F)

19  (PSI)

Viscosity

(Range)

5÷5000  (mPas)

Precision

(1 and 26 quarts/min)

±1  of value indicated after 

calibration  (%)

Repeatability

(Typical)

±0.3   (%)

Screen

Liquid crystals LCD 

Featuring:

-

5-figure partial

6-figure Reset Total plus x10 / x100

6-figure non reset Total plus x10 / x100

Power supply

(2) 1.5 V alkaline batteries size AAA

Battery life

14,000 - 100,000 hours

Weight

.83 lbs. (including batteries)

NOTE: The Reset Total and Total will be automatically 

            converted to the new unit of measurement. No 

            new calibration is required when changing the 

            unit of measurement.

!

b)

a)

NOTE: 

Refer to your local regulations before disposing  the old 

batteries.

!

1)

2)

3)

4)

c)

NOTE: Always make sure the liquid has been drained 

from the meter before cleaning

.

!

NOTE:  Only  one  of  the  two  gears  has  magnets.These 

gears must be inserted into the body of the meter, and 

placed perpendicular to each other. To position the gears 

correctly, refer to the diagram below, making sure that 

the position of the gear with respect to the magnets is 

observed.

 

If the gears are not installed correctly, the  me-

ter will not work. Check that the gears are rotating freely 

before closing the cover.

!

1)

2)

3)

4)

5)

6)

1)

2)

3)

4)

5)

4-1/4”

2.46”

3/

4”

2.

38

2.54”

1/2" NPT

3.21”

2.

55

3.

57

11

5

8

9
6
4

3

12

2

10

13

15

1

7

14

16

d)

!

NOTE: 

The gear with the magnet (TIM-615-13) must be pla-

ced so that it is on the left side of the meter when assmbled

Summary of Contents for PRO TIM-615

Page 1: ...speed of change increases the longer the button is held If you accidentally program the wrong value repeat the operation starting from step number 6 8 Saving the New Calibration Factor Before performing this operation double check to make sure the display value on the meter is the same value that is in the calibrated container Press the RESET button for at least one second The TIM 615 calculates t...

Page 2: ...ies If you continue to use the meter without changing the batteries the second battery alarm level will eventually be reached preventing meter operation In this condition the battery icon will flash and will be the only item visible on the LCD You must change the batteries once the meter has reached this mode All dispensing data from this point forward will not be displayed or stored in the meter ...

Page 3: ...letters CAL FIELD appear Dispense 4 quarts of oil into your container Start and stop the flow as needed to reach the appropriate level in the container Press and hold the RESET button until an arrow appears in the lower left hand corner of the display It will read CAL Direct with an arrow pointing up If the total is under 4 quarts use this setting If the total is over 4 quarts press and release th...

Page 4: ...trigger to purge any air from your hose into the bucket 12 When you have a steady flow of oil coming out stop dispensing and close the non drip tip on the control handle 13 Dispose of the oil in the bucket into your waste drain Trouble Shooting Control Handles There is a display problem o Have you changed the batteries The meter is dispensing oil but not metering anything o Have you changed the ba...

Page 5: ...600 FMHF 7 Flex Manual Hi Flow TIM 743 1 TIM 600 RA 4 Rigid Automatic TIM 746 TIM 600 RM 2 Rigid Manual TIM 753 TIM 600 RMHF 6 Rigid Manual Hi Flow TIM 752 These tip assemblies have had their compression nuts Loctited at the factory during the assembly process We recommend placing a few drops of Loctite 680 Retaining Compound on the threads of the compression nut when replacing these tip assemblie...

Page 6: ...ng normal use Do not under any circumstances place Loctite on the threads that are used to fasten the tip assembly to the meter body Filter cleaning procedure 1 remove swivel 2 remove clean reinstall filter 3 reinstall swivel Model Number Tip Style Description Tip Assembly TIM 601 FA 3 Flex Automatic TIM 747 TIM 601 FM 1 Flex Manual TIM 743 TIM 601 RMGO 5 Swivel 90º Rigid Manual TIM 742 TIM 601 FM...

Page 7: ...airable The TIM 715 meter can be replaced with a new TIM 615 meter You can reuse your old extension nozzle and old control handle with the TIM 615 Please call 410 252 9300 and press 3 for pricing and availability The upper of numbers reset to zero after a battery replacement and should count up as oil is pumped through the meter The number in the lower right corner is not resettable ...

Reviews: