MAINTENANCE
LAB STORAGE TANKS
23
98-0123 Rev E
9.3
UV Sensor Cleaning / Replacement
Mineral deposits and sediment may accumulate on the sensor window over time, decreasing the UV
energy detected. If the UV controller indicates that the UV intensity is low, one cause may be a stained
quartz sleeve and / or sensor window.
WARNING:
The UV sensor is an extremely sensitive and fragile instrument. Extreme care
is required when handling. The sensor window is constructed from quartz, which is
extremely fragile. Use caution when handling the sensor to ensure that you do not chip or
break this quartz window.
Prerequisites:
Remove and clean the quartz sleeve.
NOTE:
The
quartz sleeve and UV sensor should be cleaned at the same time.
Disconnect the UV sensor
from the controller by
disconnecting the sensor
cable.
Remove the UV sensor by
grasping the body of the
sensor and rotating it
counter-clockwise.
Submerge only the end of
the sensor in the
commercial scale remover
for 30 minutes.
Clean the sensor with a
cotton swab and spray
with water.
Re-insert the UV sensor
until a water-tight seal is
achieved.
Summary of Contents for 00HC-4090
Page 2: ......
Page 7: ...COMPONENT IDENTIFICATION LAB STORAGE TANKS 5 98 0123 Rev E 2 COMPONENT IDENTIFICATION...
Page 8: ...COMPONENT IDENTIFICATION LAB STORAGE TANKS 6 98 0123 Rev E...
Page 9: ...COMPONENT IDENTIFICATION LAB STORAGE TANKS 7 98 0123 Rev E 2 1 Flow Schematic...
Page 15: ...OPERATION LAB STORAGE TANKS 13 98 0123 Rev E Figure 3 Grundfos Eye Indicators...
Page 16: ...OPERATION LAB STORAGE TANKS 14 98 0123 Rev E 5 2 Default Pump Program...
Page 28: ...REPLACEMENT PARTS LAB STORAGE TANKS 26 98 0123 Rev E 11 REPLACEMENT PARTS...