48
98-0157 Rev P
6.5.4
Line Cells Screen
The Line Cell menu allows the user to access the calibrated numbers used to determine the
conductivity of the water in the unit. This should not be altered.
Single Pass Screen
Dual Pass Screen
Figure 56
Figure 57
OFFSET and GAIN are calibration parameters. Offset adds or subtracts a constant value to the signal.
Gain multiplies the signal by a constant factor.
6.5.5
RO Heat Disinfect Screen
Figure 58
The RO Heat Disinfect menu allows
access to the PUMP RUN SLOW and
COOL DOWN END TEMP.
PUMP RUN SLOW TEMP is the set-point
at which the pump on the system will run
at a lower pressure than normal. This is
required during a heat disinfection of the
system because the membranes cannot
handle high pressures with water over
113 degrees F at >45 psi of pressure.
COOL DOWN END TEMP is the
temperature at which the cool down
cycle will end after a heat disinfection of
the RO. At this temperature, the RO will
be able to supply water to the loop.
Summary of Contents for 00MDP2-230
Page 2: ......
Page 61: ...59 98 0157 Rev P 9 2 1 MSP Spare Parts ...
Page 62: ...60 98 0157 Rev P ...
Page 63: ...61 98 0157 Rev P 9 2 2 MDP Spare Parts ...
Page 64: ...62 98 0157 Rev P ...
Page 66: ...64 98 0157 Rev P FLOW DIAGRAMS MSP UNITS ...
Page 67: ...65 98 0157 Rev P MDP UNITS ...
Page 68: ...66 98 0157 Rev P WIRING SCHEMATIC MSP UNITS ...
Page 69: ...67 98 0157 Rev P ...
Page 70: ...68 98 0157 Rev P ...
Page 71: ...69 98 0157 Rev P ...
Page 72: ...70 98 0157 Rev P ...
Page 73: ...71 98 0157 Rev P ...
Page 74: ...72 98 0157 Rev P ...
Page 75: ...73 98 0157 Rev P ...
Page 76: ...74 98 0157 Rev P MDP Units ...
Page 77: ...75 98 0157 Rev P ...
Page 78: ...76 98 0157 Rev P ...
Page 79: ...77 98 0157 Rev P ...
Page 80: ...78 98 0157 Rev P ...
Page 81: ...79 98 0157 Rev P FX3U 4AD ADP ...
Page 82: ...80 98 0157 Rev P ...
Page 83: ...81 98 0157 Rev P ...
Page 86: ...84 98 0157 Rev P ...
Page 87: ...85 98 0157 Rev P MDP Units ...
Page 88: ...86 98 0157 Rev P ...