55
98-0157 Rev P
For hard water areas even on systems operating on a softened water supply it is recommended that
the unit is chemically cleaned using AmeriClean A.
For soft water areas, cleaning using AmeriClean B, organic removal formulation is recommended in
place of AmeriClean A.
WARNING:
Failure to test sample may lead to chemicals entering the distribution loop.
DANGER:
Please read the appropriate material safety data sheets and always wear the
recommended personnel protective equipment before handling any of the chemicals.
CAUTION:
(i) Warn all end users that chemical cleaning is taking place.
(ii) Stop dialysis, check with Renal technicians.
(iii) Isolate distribution loop main or other systems by closing appropriate
manual isolation valves.
WARNING:
Wear personal protective equipment when conducting the chemical clean process
EYE WEAR, CLOTHING, GLOVES.
Method:
Test the pH of the water before beginning to know the end point of the cleaning. When the water
returns to this pH that means the chemical has been rinsed out.
Chemical cleaning of the unit can be initiated from the OPERATION MENU of the touch screen.
Stop the MediQA by pressing the OFF button on the OPERATION MENU.
Depress CHEMICAL CLEAN button.
Wait for prompt to ADD CLEANING CHEMICAL. Remove the lid of the tank and add chemical to tank.
Slowly add the chemical in the amount listed in the above table.
IF USING POWDER, AGITATE THE
SOLUTION UNTIL ALL OF THE POWDER HAS DISSOLVED PRIOR TO ADDING TO BREAK TANK
.
Replace the tank lid.
Depress ADD CLEANING CHEMICAL button. Once added, the cleaning chemical will be automatically
circulated around the MediQA.
On completion of the automatic circulation and drain down / flush, a prompt TEST FOR RESIDUE
CHEMICALS is displayed.
A permeate water sample
must
be taken and tested for any presence of cleaning chemical. Check
with renal clinical / technicians that water quality is acceptable for use.
If acceptable for use depress TEST FOR RESIDUE CHEMICALS once the sample is found to be
satisfactory.
Summary of Contents for 00MDP2-230
Page 2: ......
Page 61: ...59 98 0157 Rev P 9 2 1 MSP Spare Parts ...
Page 62: ...60 98 0157 Rev P ...
Page 63: ...61 98 0157 Rev P 9 2 2 MDP Spare Parts ...
Page 64: ...62 98 0157 Rev P ...
Page 66: ...64 98 0157 Rev P FLOW DIAGRAMS MSP UNITS ...
Page 67: ...65 98 0157 Rev P MDP UNITS ...
Page 68: ...66 98 0157 Rev P WIRING SCHEMATIC MSP UNITS ...
Page 69: ...67 98 0157 Rev P ...
Page 70: ...68 98 0157 Rev P ...
Page 71: ...69 98 0157 Rev P ...
Page 72: ...70 98 0157 Rev P ...
Page 73: ...71 98 0157 Rev P ...
Page 74: ...72 98 0157 Rev P ...
Page 75: ...73 98 0157 Rev P ...
Page 76: ...74 98 0157 Rev P MDP Units ...
Page 77: ...75 98 0157 Rev P ...
Page 78: ...76 98 0157 Rev P ...
Page 79: ...77 98 0157 Rev P ...
Page 80: ...78 98 0157 Rev P ...
Page 81: ...79 98 0157 Rev P FX3U 4AD ADP ...
Page 82: ...80 98 0157 Rev P ...
Page 83: ...81 98 0157 Rev P ...
Page 86: ...84 98 0157 Rev P ...
Page 87: ...85 98 0157 Rev P MDP Units ...
Page 88: ...86 98 0157 Rev P ...