58
98-0157 Rev P
MAINTENANCE
There are no user replaceable parts on the MediQA.
The MediQA is a medical device; as such modifications to the device are not permitted.
Modifications to the machine by anyone other than AmeriWater personnel will invalidate
marketing clearance for the device.
For all service enquiries contact AmeriWater at:
Tel No: 1-937-461-8833
9.1
Thermostat Testing
AmeriWater recommends the thermostat on the heating element be tested annually by qualified
personnel. To conduct this testing, access the device when no patients are under treatment.
To begin, remove the lid from the MediQA storage tank. Select the calibration menu and log in (see
Section 6.5). Access the OVERRIDES menu and open the fill solenoid valve. Allow the tank to fill to
between 50-75% and close the fill valve. Ensure that the heater is completely submerged.
Place a known good thermometer into the tank away from the heater. Turn on the heater on the
OVERRIDES page.
CAUTION:
During testing, water temperature will be sufficient to create steam, which is hot
enough to scald. Ensure the temperature can be determined without reaching into the tank.
Verify that the heater shuts off between 185 – 210° F.
If the thermometer on the heater does not shut off at the prescribed temperature, contact
AmeriWater for guidance.
Navigate to the OVERRIDES screen and open the fill solenoid valve and allow the tank to completely
fill. Close the fill solenoid. Doing this will reduce the temperature of the water. Open the drain valve
and allow the tank to completely drain. Close the drain valve and remove the thermometer from the
tank and refill the tank before replacing the lid.
9.2
Spare Parts
Should a faulty item be identified, please notify our service department of the item part number.
The MediQA is classified as a medical device. Replacement parts should be recorded using the log
sheet in Section 9.3. This must be done in order to maintain the integrity of the machine throughout
its operational life in accordance with medical device regulations.
Failure to install AmeriWater supplied replacement parts will invalidate the warranty. If replacement
parts are not installed, then the responsibility for any subsequent incidents linked to the replaced
part or affected parts rest with the buyer / hospital / maintenance staff or service provider.
Summary of Contents for 00MDP2-230
Page 2: ......
Page 61: ...59 98 0157 Rev P 9 2 1 MSP Spare Parts ...
Page 62: ...60 98 0157 Rev P ...
Page 63: ...61 98 0157 Rev P 9 2 2 MDP Spare Parts ...
Page 64: ...62 98 0157 Rev P ...
Page 66: ...64 98 0157 Rev P FLOW DIAGRAMS MSP UNITS ...
Page 67: ...65 98 0157 Rev P MDP UNITS ...
Page 68: ...66 98 0157 Rev P WIRING SCHEMATIC MSP UNITS ...
Page 69: ...67 98 0157 Rev P ...
Page 70: ...68 98 0157 Rev P ...
Page 71: ...69 98 0157 Rev P ...
Page 72: ...70 98 0157 Rev P ...
Page 73: ...71 98 0157 Rev P ...
Page 74: ...72 98 0157 Rev P ...
Page 75: ...73 98 0157 Rev P ...
Page 76: ...74 98 0157 Rev P MDP Units ...
Page 77: ...75 98 0157 Rev P ...
Page 78: ...76 98 0157 Rev P ...
Page 79: ...77 98 0157 Rev P ...
Page 80: ...78 98 0157 Rev P ...
Page 81: ...79 98 0157 Rev P FX3U 4AD ADP ...
Page 82: ...80 98 0157 Rev P ...
Page 83: ...81 98 0157 Rev P ...
Page 86: ...84 98 0157 Rev P ...
Page 87: ...85 98 0157 Rev P MDP Units ...
Page 88: ...86 98 0157 Rev P ...