2  

2.0  THEORY OF OPERATION 

Mix Tank 

The mix tank is filled to a desired level with RO purified water.  The tank filling process is done 
MANUALLY, and is restricted by a 3 GPM flow control.  As the bicarbonate powder is added, 
the system utilizes the pump discharge pressure, forcing the liquid through an eductor to mix 
and quickly dissolve the powder. 

Disinfection 

The mix tank is designed for quick disinfection. The mix tank utilizes a spray nozzle design to 
completely spray the inside of the tank during the disinfection process.   

Summary of Contents for Bicarb

Page 1: ...Bicarb Manual Mix and Dispensing Installation Operation Manual Manufactured With Pride In The USA www ameriwater com 800 535 5585 AmeriWater 3345 Stop 8 Rd Dayton OH 45414 P N 98 0134 REV C ...

Page 2: ...UCTION 1 2 0 THEORY OF OPERATION 2 3 0 INSTALLATION 3 3 1 INSTALLATION REQUIREMENTS 3 4 0 START UP 4 4 1 DISINFECTION WITH CHLORINE OR PERACETIC ACID 5 4 2 MIXING BICARB 7 5 0 MAINTENANCE 8 6 0 SYSTEM DRAWINGS 9 7 0 PARTS LIST 12 ...

Page 3: ...carbonate powder in solution The Bicarb System is designed to mix and dispense Bicarb solutions The system is not equipped with monitoring devices such as conductivity or pH Therefore the solution manufacture s procedures should be followed and are the sole responsibility of the user CAUTION Federal law restricts this device to sale by or on the order of a physician for use in hemodialysis applica...

Page 4: ...y a 3 GPM flow control As the bicarbonate powder is added the system utilizes the pump discharge pressure forcing the liquid through an eductor to mix and quickly dissolve the powder Disinfection The mix tank is designed for quick disinfection The mix tank utilizes a spray nozzle design to completely spray the inside of the tank during the disinfection process ...

Page 5: ...er installation and operation of the Bicarb Manual Mix and Dispensing System 1 Locate the Bicarb System on a level floor 2 Locate the system as close as possible to a floor drain to facilitate draining of the system 3 Connect the purified water supply to the inlet hose located at the top of the pump header using the inside diameter medical grade hose provided 4 Plug the Bicarb Pump Control into a ...

Page 6: ...eding with the start up instructions Refer below Fig 4 1 to reference control panel and valve designations WARNING The Bicarb System must be disinfected prior to placing the system in service Failure to comply may result in injury illness or death to the patients Fig 4 1 BICARB CONTROL PANEL AND VALVE DESIGNATIONS ...

Page 7: ...he Bicarb System are closed 3 Open the RO water supply valve and WATER INLET valve 4 Open the valve marked 3M 5 Allow RO purified water to fill the mix tank to the desired level a 55 gallon mix tank Fill to 25 gallons and close WATER INLET valve b 100 gallon mix tank Fill to 50 gallons and close WATER INLET valve 6 Once filled close the 3M and RO water supply valve then open valve 2M NOTE If 3M va...

Page 8: ... and close WATER INLET valve b 100 gallon mix tank Fill to 50 gallons and close WATER INLET valve 18 Turn the MIX OFF ON switch to ON 19 Rinse tank piping by opening valve 2M and closing valve 3M for several minutes 20 Open valve 3M and close valve 2M for several minutes 21 Once the rinse water is thoroughly circulated in the tank and has sprayed the tank s inner surfaces open valve 1D Then the di...

Page 9: ...LET WATER valves to fill the mix tank 6 Open valve 3M and manually fill tank to desired level 7 Once Mix tank is filled to the desired level close valve 3M and RO water supply valve 8 Open valve 2M to send water to the mix nozzle eductor 9 Turn the MIX OFF ON switch to ON and open the tank lid 10 Slowly add the dry powder to the mixing water and close the tank lid 11 After the mix time is complete...

Page 10: ... the hoses are free of restrictions that may reduce the flow of the Bicarb System Weekly Maintenance AmeriWater recommends disinfecting the Bicarb System weekly or as directed by your facilities protocol Quarterly Maintenance AmeriWater recommends that the Bicarb System be decalcified quarterly 4 times per year Vinegar is an acceptable solution for decalcifying the system Your Bicarb System may ha...

Page 11: ...9 6 0 SYSTEM DRAWINGS ...

Page 12: ...10 ...

Page 13: ...11 ...

Page 14: ... TYGON HY PURITY SILICONE TUBING 12677125 12 0011 RELAY 30AMP 300V SPST NO 120V COIL 61 0002 PUMP 1ST 5HP 115V ODP 80 0102 SWITCH 2 POSITION MAINTAIN 65511221 VALVE BALL 1 2 TRUE UNION PVC80 VALVE BALL 1 2 TRUE UNION CPVC80 041531812 041 0125 VALVE NOZZLE 5HB DISPENSING POLYPRO 041 0041 VALVE CHECK 1 2 FPT WHITE PP VALVE CHECK BALL 3 4 PVC80 55 0003 041831809 VALVE BALL EL 38 MPTx 25FPT PVC80 0410...