Cool down period
Once the water availability stage has timed out the system will
proceed into automatic “Cool down”. The system will perform a
series of tank drain downs and introduce cold RO permeate into the
distribution loops until the return temperature on the system falls
below the “maximum service temperature” setting. The factory
default for this can be found in Section 6.10. On reaching this
temperature the system automatically disconnects itself from the
loop(s) and re-introduces the RO unit(s).
Drain in Operation
Following the Cool down period the tank drain valve will remain
open until the tank level falls to zero at which point the unit will
return to
STANDBY.
To ensure that the tank remains empty during
STANDBY
the drain valve will open
for 5 minutes every 2 hours.
Standby
STANDBY
will be displayed to indicate that the unit is “healthy” and
waiting for the next heat disinfection routine to initiate.
Remote Light
A four tier remote light is available to clearly and easily show the
operational state of the Heatsan.
Light Pattern
Light
Light
Pigtail
Wire
Termination
Heatsan
PLC
Output
Illumination
Description
Pattern
Time
Green
Green
K2 (12)
Y21
System is in
standby mode
Steady
System is in drain
in operation mode
1 sec on
1 sec off
Amber
Orange
K3 (14)
Y22
System is in tank
preheat
Steady
Blue
Blue
K4 (14)
Y23
System is preparing
to send hot water
to distribution loop
1 sec on
1 sec off
Hot water is
circulating in
distribution loop
Steady
Red
Red
KAB (14)
Y6
General Alarm
condition
Steady
Common
Yellow
TB1, 24C
N/A
N/A
N/A
N/A
Summary of Contents for Heatsan 00HS208
Page 51: ...50 98 0156 Rev G Heatsan Process Flow Diagram ...
Page 52: ...51 98 0156 Rev G Wiring Diagrams ...
Page 53: ...52 98 0156 Rev G ...
Page 54: ...53 98 0156 Rev G ...
Page 55: ...54 98 0156 Rev G ...
Page 56: ...55 98 0156 Rev G ...
Page 57: ...56 98 0156 Rev G ...
Page 58: ...57 98 0156 Rev G Installation Layout Water Connection Detail ...
Page 62: ...61 98 0156 Rev G Tubing with Heatsan and MediQA Hidden ...
Page 64: ...63 98 0156 Rev G RO Divert Valve and Heatsan Divert Valve 1 Heatsan Divert Valve 2 ...
Page 65: ...64 98 0156 Rev G Heatsan Discharge Solenoid SV3A NC Heatsan Fill Solenoid SV3B NO ...
Page 70: ...69 98 0156 Rev G ...