65
98-0156 Rev G
Once all plumbing has been connected, run the cable for the solenoid
valves toward the panel on the Heatsan.
The NC solenoid valve on the HS DISCHARGE line indicated above is
labeled as “SV3A” and should be connected to the “SV3A” port on Heatsan
panel.
The NO solenoid valve on the HS FILL line indicated above is labeled as
“SV3B” and should be connected to the “SV3B” port on Heatsan panel.
The remaining sanitary connections are for the Heatsan return, drain and
overflow connections.
Mount the control for the interface between the MROZ and Heatsan close
to the control valves. Ensure that the cables for the valves will be able to
reach the control panel (25’ maximum cable length).
Connect the cables to the following connectors: VC1 (RO Divert Valve) to
VC1, VC2 (Heatsan Divert Valve #1) to VC2 and VC3 (Heatsan Divert Valve
#2) to VC3.
Run cable ICZ to the control panel interface connector ICZ. Terminate the
flying leads into the MROZ side entry hood at the tank full high (terminals
1 & 6) and tank full low (terminals 2 & 7). Conductor 1 to terminal 1,
conductor 2 to terminal 6, conductor 3 to terminal 2 and conductor 4 to
terminal 7.
Run cable ZINT from the Heatsan connector ZINT to the control panel
interface connector ZINT.
Plug the control panel in to an 115V outlet.
Summary of Contents for Heatsan 00HS208
Page 51: ...50 98 0156 Rev G Heatsan Process Flow Diagram ...
Page 52: ...51 98 0156 Rev G Wiring Diagrams ...
Page 53: ...52 98 0156 Rev G ...
Page 54: ...53 98 0156 Rev G ...
Page 55: ...54 98 0156 Rev G ...
Page 56: ...55 98 0156 Rev G ...
Page 57: ...56 98 0156 Rev G ...
Page 58: ...57 98 0156 Rev G Installation Layout Water Connection Detail ...
Page 62: ...61 98 0156 Rev G Tubing with Heatsan and MediQA Hidden ...
Page 64: ...63 98 0156 Rev G RO Divert Valve and Heatsan Divert Valve 1 Heatsan Divert Valve 2 ...
Page 65: ...64 98 0156 Rev G Heatsan Discharge Solenoid SV3A NC Heatsan Fill Solenoid SV3B NO ...
Page 70: ...69 98 0156 Rev G ...