DISINFECTION AND CLEANING
Advisory Note
The
Heatsan
unit during normal operation will not require being
chemically cleaned or disinfected.
The heat disinfection process will also serve to disinfect the unit’s
integral tank and internal pipework.
Over a period of time (12 months) scale deposits could build up on the
internal tank walls and heater elements. De-scaling of the unit must
only be carried out by AmeriWater or under instruction by AmeriWater.
Chemical cleaning of the unit by un-trained personnel or the use of un-
approved chemicals is strictly forbidden and may invalidate any
guarantees or warranty.
It is recommended that the tank and heaters are inspected annually
for signs of fouling/scaling and general integrity.
If the unit is to be decommissioned and placed in long term storage, >6
months, it should be fully drained, i.e. drain the pipework and
recirculation pump fully and isolate its feedwater supply. DO NOT fill
the unit with any form of preservative.
Surface Cleaning
When cleaning the exterior surfaces of the device, it is recommended
that you use a soft non-marking cloth dampened with water. Do not
use chemical cleaning agents. If it is necessary to use a chemical
cleaning agent, contact AmeriWater for permission prior to use.
Summary of Contents for Heatsan 00HS208
Page 51: ...50 98 0156 Rev G Heatsan Process Flow Diagram ...
Page 52: ...51 98 0156 Rev G Wiring Diagrams ...
Page 53: ...52 98 0156 Rev G ...
Page 54: ...53 98 0156 Rev G ...
Page 55: ...54 98 0156 Rev G ...
Page 56: ...55 98 0156 Rev G ...
Page 57: ...56 98 0156 Rev G ...
Page 58: ...57 98 0156 Rev G Installation Layout Water Connection Detail ...
Page 62: ...61 98 0156 Rev G Tubing with Heatsan and MediQA Hidden ...
Page 64: ...63 98 0156 Rev G RO Divert Valve and Heatsan Divert Valve 1 Heatsan Divert Valve 2 ...
Page 65: ...64 98 0156 Rev G Heatsan Discharge Solenoid SV3A NC Heatsan Fill Solenoid SV3B NO ...
Page 70: ...69 98 0156 Rev G ...