MAINTENANCE
Maintenance
The
Heatsan
unit does not contain any consumables that require
replacement on a regular basis.
Only approved spare parts provided by AmeriWater are to be
used. If un-authorized spare parts are supplied or fitted this
could invalidate the warranty guarantee, affect the unit’s
performance or compromise the safe operation of the unit.
Maintenance on the unit must only be carried out by AmeriWater or by an
approved/authorized third party. Routine maintenance by the operator is
not required. It is, however, recommended that should any fault occur
with the unit that it is reported to the appropriate party responsible for
maintaining the water treatment system.
Thermostat Testing
AmeriWater recommends that the thermostats on the heating elements
be tested annually by qualified personnel. To conduct this testing, you will
need access to a loop with no patients under treatment.
CAUTION:
During this testing, the temperature of the water will
be sufficient to create steam, which is hot enough to scald
during this test. Ensure that you will be able to determine the
temperature without reaching into the tank.
To begin, remove the lid from the Heatsan system. Log in as user level # 3
(see Section 6.2). Access the manual operation menu and open the fill
solenoid valve. Allow the unit to fill to 16 gallons of water and close the fill
valve.
Using a calibrated thermometer, insert this into the tank away from the
heaters. Use caution once the heater has turned on as the water will
become hot enough to scald. Turn on the recirculation pump as well as
Heater #1. Verify that this heater (2K1M) shuts off between 185 – 210° F.
Turn off Heater #1. Open the fill valve and allow the tank to fill to 24
gallons. The added water will lower the temperature in the tank. Close the
fill valve and turn on Heater #2. Verify that this heater (3K1M) shuts off
between 185 – 210° F.
Turn off Heater #2. Open the fill valve and allow the tank to fill to 32
gallons. Close the fill valve and turn on Heater #3. Verify that this heater
(4K1M) shuts off between 185 – 210° F.
Summary of Contents for Heatsan 00HS208
Page 51: ...50 98 0156 Rev G Heatsan Process Flow Diagram ...
Page 52: ...51 98 0156 Rev G Wiring Diagrams ...
Page 53: ...52 98 0156 Rev G ...
Page 54: ...53 98 0156 Rev G ...
Page 55: ...54 98 0156 Rev G ...
Page 56: ...55 98 0156 Rev G ...
Page 57: ...56 98 0156 Rev G ...
Page 58: ...57 98 0156 Rev G Installation Layout Water Connection Detail ...
Page 62: ...61 98 0156 Rev G Tubing with Heatsan and MediQA Hidden ...
Page 64: ...63 98 0156 Rev G RO Divert Valve and Heatsan Divert Valve 1 Heatsan Divert Valve 2 ...
Page 65: ...64 98 0156 Rev G Heatsan Discharge Solenoid SV3A NC Heatsan Fill Solenoid SV3B NO ...
Page 70: ...69 98 0156 Rev G ...