Page 37
User Guide
Signal Processor
LMG AE
3.5 Configuring for ABT/S Thermometers
Enter the processor menu by pressing the (
ENTER
) key.
Set the entry code using the (
UP
), (
DOWN
) keys to 417 and press (
ENTER
).
Now select the channel to be configured for Snap Shot operation.
Enter the channel Configuration menu.
Configure the channel to the required settings ensuring that ABT/S
is selected.
Configure any other channels as required, when complete select the
Exit
option and
Save
the settings.
Using the (
UP
), (
DOWN
) keys select the system Configuration menu and
press (
ENTER
).
Now select the channel previously configured for ABT/S.
Enter that channel and select the Snap Shot display option.
The following parameters will need to be configured for Snap Shot:
(a) Display range Tmax and Tmin.
(b) Interval time.
(c) Threshold level.
(d) Pre trigger value.
(e) Position of cursor 1.
(f) Position of cursor 2.
Having completed the above configuration menu select the
Exit
option and
Save
the settings.
If multiple thermometers with a multi channel processor are being used,
then a multi channel display may be required. If so, the multi channel display
menu will need to be used.
Using the (
UP
), (
DOWN
) keys select the multi channel display option.
On entry in to the multi channel display menu ensure that the display type is
set to
MUL
SNAP
SHOT
, using the (
UP
), (
DOWN
) keys to select the option.