User Guide
5 - 6
Signal Processor
LMG MkII
Name
Menu title text
Range
Increment
Default
Comments
Min
Max
Units Menu value text
MT2 Maths Function 2 Type
0
1
1
0
0 = Max
1 = Min
2 = Mean
3 = Diff
4 = Range
5 = BGComp
MC2 Maths function 2
Allocation
Add together:
ChA = 128
ChB = 64
ChC = 32
ChD = 16
MH2 Maths Function 2
Tmax
1
Dependent upon
thermometer
ML2 Maths Function 2
Tmin
1
Dependent upon
thermometer
MA2 Maths Function 2
Alarm 1 Level
1
Dependent upon
thermometer
MM2 Maths Function 2
Alarm 1 Mode
0
1
1
0
MO2 Maths Function 2
Output Mode
0
1
1
1
mA
0 = 0 to 20mA
1 = 4 to 20mA
ST2
Maths Function 2
Temperature
System output
value. Read only
SS2
Maths Function 2
Status
0
255
Bits in the Status
byte:
0 = Not set
1 = Set
2 = Alarm A2
3 = Error
4 = Over range
5 = Under range
6 = Units
7 = Alarm A1
SE1
F1 emiss value for BG
comp
300
1000
1
1000
300 = 0.300
1000 = 1.000
OS1 F1 offset correction
value for BG
0
100
1
0
deg
Offset factor for
BG compensation
SE2
F2 emiss value for BG
comp
300
1000
1
1000
300 = 0.300
1000 = 1.000
OS2 F2 offset correction
value for BG
0
100
1
0
deg
Offset factor for
BG compensation
Summary of Contents for LMG MkII
Page 6: ...INTRODUCTION 1 ...
Page 10: ...INSTALLING THE PROCESSOR 2 ...
Page 14: ...User Guide 2 4 Signal Processor LMG MkII Fig 2 4 LMG MkII Electrical System Overview ...
Page 22: ...3 USING THE PROCESSOR ...
Page 44: ...TIME FUNCTION PROCESSING 4 ...
Page 50: ...SERIAL COMMUNICATIONS 5 ...
Page 62: ...Signal Processor LMG MkII Blank ...
Page 63: ...EXTERNAL I O MATHS FUNCTIONS 6 ...
Page 64: ...Signal Processor LMG MkII Screws A Fig 6 1 Location of screws on back panel Screws A ...
Page 88: ...MAINTENANCE 7 ...
Page 92: ...User Guide 7 4 Signal Processor LMG MkII Fig 7 3 USB Export Data option ...