55
A01-00071 Rev F
www.powervar.com
A03-00410 Rev. C
www.powervar.com
Security II Rackmount UPM Specifications
International 1440 VA Models - Medical
Front Panel Controls
• Power On/Off
• Test
• Load Level LED Gauge
• Battery Charge LED Gauge
• Voltage Manager Boost LED
• Voltage Output On LED
• Voltage Manager Buck LED
• On Battery LED
• Replace Battery LED
• Overload LED
• Fault Code LED
• System Over Temperature LED
Rear Panel Information and Controls
• 6 Foot Power Cord with IEC320-C14 Plug
• Integrated Switched Power Distribution Unit
(Requires SNMP / Web Card)
•
Configuration Manager DIP Switches
• Communications Manager DB9 Port, USB Port
• Circuit Breaker (AC and DC)
• AC Inlet Module
• Communication Option Slot: SNMP / Web Card, Isolated
Relay Contacts Card, or Remote Off Card
• DC External Battery Cabinet Connector
• Fan (Forced Cooling)
Internal Batteries
• User Hot-Swappable (See Instruction Manual)
• Type: 12 Volt, High Rate 23W
• Quantity: 4 Batteries
• Recharge Time: 6 Hours to 80%, 24 Hours to Full Charge
(For Internal Batteries Only)
Environmental
•
Temperature:
0 - +40℃(32 to 104℉) Operating
-20 to 60℃(-40 to 140℉) Shipment/
Storage
• Humidity: 5 to 90% Non-Condensing (Operating,
Shipment/Storage)
• Altitude: 3,000m (10,000 ft) max. Operating;
12,000m (40,000 ft) max. Shipment/Storage
Safety Agency and EMC Compliance:
All Units are Listed by UL and Marked With the UL/cUL Marking
Standard UPM:
Products Listed to:
• UL, IEC62042, CE Mark with CB Report
Products in Compliance with:
• FCC-Part 15, Subpart B, Sections 15.107b & 15.109b
• Class A Digital Device*
• CISPR11:2009, A1; 2010, Class A*
• IEC61000-4-2, Electrostatic Discharge
• IEC61000-4-3, Radiated Electromagnetic Field Immunity
• IEC61000-4-4, Electrical Fast Transient/Burst Immunity
• IEC61000-4-5, Surge Immunity
• IEC61000-4-6, Immunity to Conducted Radio Frequency
• Disturbances
• IEC61000-4-8, Power Frequency Magnetic Field Immunity
• IEC61000-4-11, Voltage Dips, Short Interruptions, and
• Voltage Variations
*Note: Class B is Available as an Option
Please Consult Your Powervar Sales Representative
RoHS Compliance:
All Products (Standard and Medical) are RoHS Compliant
1440 VA MODELS TYPICAL RUN-TIMES (MINs)
10% (130W) 20% (259W)
30% (389W) 40% (518W) 50% (648W) 60% (778W) 70% (907W) 80% (1036W) 90% (1166W) 100% (1296W)
Internal Batteries Only
90
38
20
12
9
6
5
4
2.8
2.5
In 1) E4804-12
480
183
110
72
59
51
35
30
25.8
24.5
In 2) E4804-12
870
328
200
132
109
96
65
56
48.8
46.5
In 3) E4804-12
1260
473
290
192
159
141
95
82
71.8
68.5
In 4) E4804-12
1650
618
380
252
209
186
125
108
94.8
90.5
In 5) E4804-12
2040
763
470
312
259
231
155
134
117.8
112.5
In 6) E4804-12
2430
908
560
372
309
276
185
160
140.8
134.5
In 7) E4804-12
2820
1053
650
432
359
321
215
186
163.8
156.5
In 8) E4804-12
3210
1198
740
492
409
366
245
212
186.8
178.5
NOISE REJECTION-ISOLATION:
With unit under power and an ANSI/IEEE C62.41Cat. A pulse applied either normal or common mode at the input, the noise
output voltage will be less than 10V normal mode and less than 0.5V common mode in all four quadrants (CM-NM, NM-NM, CM-CM, NM-CM).
SURGE VOLTAGE WITHSTAND CAPABILITY:
Tested under power to ANSI/IEEE C62.41 Cat. A & B (formerly IEEE587-1980). Cat. A - 6000V @ 200 amps,
Cat. B - 6000V @ 500 amps..
Warranty/Support:
Powervar warrants the electronics and transformers used in its uninterruptible power managers to be free from defects in materials and work-
manship for a period of five years from the date of shipment. Batteries are warranted for a period of two years from the date of shipment. For service or support on
any Powervar product, please contact Powervar Technical Support at (800) 369-7179 or visit the Powervar website at
www.powervar.com
.
Battery Life Disclaimer: Powervar’s standard battery warranty applies only to UPS and UPM products which are continuously connected to AC mains power,
except during utility power outages. Products which are regularly and intentionally disconnected from AC mains power will experience battery discharge/charge
cycles potentially far more numerous than those for which the battery was designed. As a result, products used in such applications will experience substantially
reduced battery life. For that reason, Powervar’s standard battery warranty does not apply for applications in which the UPS or UPM product is regularly and
intentionally disconnected from AC mains power. Powervar UPS and UPM products used in such applications shall receive a 90 day warranty on batteries.
Security II Rackmount UPM Specifications
International 1440 VA Models - Medical