User Manual
252
i Series / iX Series
This manual is
not
intended to elaborate on the intent or purpose of the immunity tests and the
expected behavior of the EUT as described in the MIL-STD-704 documents. It is assumed that
the end-user is familiar with the content of the MIL-STD-704 test standard.
9.8.2.2 Test Coverage
The comprehensiveness of the test that can be performed with the
–704 option is determined
primarily by the capabilities of the power source used. Tests that are outside the hardware
capabilities of the power source used will not be enabled and cannot be selected. For these
tests, additional equipment may be required as indicated. The extent of coverage of the
–704
option as implemented on the IX Series platform is shown in the table below. Tests marked with
an „x‟ generally require additional hardware to be used. Tests marked „N/A‟ are not applicable to
the corresponding table. Tests marked with a „ ‟ are covered by at least one rev of the MIL-STD-
704.
No.
Description
Table
SAC
Table
TAC
Table
SVF
Table
TVF
Table
SXF
No.
Description
Table
LDC
Table
HDC
101
Load and Current
Harmonic
Measurements
101
Load Measurements
102
Steady State Limits
for Voltage and
Frequency
102
Steady State Limits for
Voltage
103
Voltage Phase
Difference
N/A
N/A
N/A
103
Voltage Distortion
Spectrum
X
X
104
Voltage Modulation
104
Total Ripple
X
X
105
Frequency Modulation
105
Normal Voltage
Transients
1
106
Voltage Distortion
Spectrum
X
X
X
X
X
201
Power Interrupt
107
Total Voltage
Distortion
301
Abnormal Steady State
Limits for Voltage
108
DC Voltage
Component
302
Abnormal Voltage
Transients
2
109
Normal Voltage
Transients
401
Emergency Limits for
Voltage
110
Normal Frequency
Transients
501
Starting Voltage
Transients
201
Power Interrupt
601
Power Failure
301
Abnormal Steady
State Limits for
Voltage and Freq.
602
Polarity Reversal
302
Abnormal Voltage
Transients
303
Abnormal Frequency
Transients
401
Emergency Steady
State Limits for
Voltage and Freq.
501
Starting Voltage
N/A
N/A
N/A
N/A
N/A
1
Note: Some restrictions apply due to voltage range limitations. See section 9.8.2.3 for details.
2
Note: Some restrictions apply due to voltage range limitations. See section 9.8.2.3 for details.
Summary of Contents for 10001i
Page 2: ......
Page 3: ......
Page 6: ...ii This page intentionally left blank...
Page 38: ...User Manual 24 i Series iX Series Figure 3 5 Rear Panel View for the 3001i 3001iX...
Page 39: ...User Manual i Series iX Series 25 Figure 3 6 Rear Panel View for the 5001i 5001iX...
Page 43: ...User Manual i Series iX Series 29 Figure 3 8 Functional Test Setup...
Page 44: ...User Manual 30 i Series iX Series Figure 3 9 Single Phase 10000 VA System 10001iX i...
Page 46: ...User Manual 32 i Series iX Series Figure 3 11 Single Phase 15000 VA System 15001iX i...
Page 48: ...User Manual 34 i Series iX Series Figure 3 13 Connection With MODE Option...
Page 118: ...User Manual 104 i Series iX Series Figure 5 2 Power Source Module Block Diagram...
Page 121: ...User Manual i Series iX Series 107 Figure 5 3 5001i Internal Layout...
Page 122: ...User Manual 108 i Series iX Series Figure 5 4 Logic Board LED s...
Page 124: ...User Manual 110 i Series iX Series Figure 5 5 AC Power Stage Layout...
Page 125: ...User Manual i Series iX Series 111 Figure 5 6 AC Control Logic Block Diagram...
Page 138: ...User Manual 124 i Series iX Series Figure 6 3 Adjustment Location...
Page 152: ...User Manual 138 i Series iX Series Figure 9 4 Voltage Modulation...
Page 219: ...User Manual i Series iX Series 205 Figure 9 36 Example Connection With 5001iX and EOS 1...
Page 221: ...User Manual i Series iX Series 207 Figure 9 38 15003iX CTS EOS3 LR3...
Page 222: ...User Manual 208 i Series iX Series Figure 9 39 15003iX 3 EOS3...
Page 233: ...User Manual i Series iX Series 219 Figure 9 41 Example Connection With MODE iX...
Page 240: ...User Manual 226 i Series iX Series Figure 9 42 Example Connections With OMNI 1 18i...
Page 241: ...User Manual i Series iX Series 227 Figure 9 43 Example Connections With OMNI 3 18i...