L-Series and FCS-18 Graphical User Interface User Manual
7.2.8 Voltage and Frequency sweep transient.
The VF Sweep transient is a compound transient that combines a Voltage
and a Frequency sweep into a single register step. When executed, both
voltage and frequency change simultaneously. Refer to section 7.2.4 and
section 7.2.7 for details on separate voltage and frequency sweeps.
The illustration shown here reflects an example of a typical VF Sweep
transient. The required row in the data entry screen for this transient would
be as shown below. In this case, register 0 was used to store the transient
information.
This transient can be used for many different types of combined
transients by judicious use of the “Goto” and “Final” values.
45 V
rms
70 ms
135 V
rms
Voltage and Frequency transient from 135 V to 45 V and 50 Hz to 300 Hz executed
over a 70 ms period.
Rev D – March 2011
California Instruments
59