User Manual – Rev J
CS Series
California Instruments
50
4.1.3
Status Indicator Lights
Eight yellow LED status indicators are located on the left hand side of the keyboard/display
panel. These LED’s correspond to the following conditions:
Hi Range
The
Hi Range
LED is on when the high voltage output range
has been selected.
Overtemp
The
Overtemp
LED indicates an overheating problem inside
the unit. This is an abnormal condition, which will cause the unit
to shut off. Check the air openings to make sure they are not
blocked.
Overvoltage
The
Overvoltage
LED indicates an output overvoltage
condition. This condition can be controlled by setting the
compliance voltage value in the PROGRAM menu. The output
will fault in the CC mode or reduce the current to limit the
voltage in the CV mode.
Remote
The
Remote
LED indicates that the unit is in remote control
mode. If the IEEE-488 interface is used, this indicator will be lit
whenever the ATN line (Attention) line is asserted by the IEEE
controller. If the RS232C, USB or LAN interface is used, the
REMOTE state can be enabled by the controller using the
SYST:REM command. Any time the
Remote
LED is lit, the front
panel of the power source is disabled. Note: The BACK button
doubles as a GOTO LOCAL button (LOCAL) while the unit is in
remote state. This allows the user to regain control of the front
panel. The LOCAL button can be disabled by sending a Local
Lockout bus command. This prevents unauthorized changes of
settings in ATE applications.
Output on/off
The
Output on/off
LED indicates the output relay status. If the
LED is off, the output relays for all output phases are open and
the external sense lines are wrapped around internally. If the
LED is on, the output relays are closed and the external sense
lines are connected to the SENSE terminal block.
ØA, ØB, ØC
The
ØA, ØB, ØC
LED indicates the output phase selection for
either settings or measurements. Phase selection can be
changed using the
Phase
button to the right of the MEMU keys.
Pressing the
Phase
button repeatedly will cycle through phase
A, B, C and ALL PHASES. The
ØA, ØB, ØC
indicators
correspond to outputs ØA, ØB and ØC on the CS Series rear
panel. If all three phase selection LED's are lit, all three phases
are selected and changes made to voltage and current limit
settings will apply to all three phases.
Note that in single-phase mode, the ØA LED is always lit and
the Phase button is inactive.