RP Series
March 2011
55
14.2.2. Source Subsystem
This subsystem programs all the output parameters of the RP Series AC source.
Subsystem Syntax
[SOURce:]
CURRent:
Set the rms current limit in amps.
VOLTage:
[LEVel]
Set the rms output voltage value in volts.
RANGe:
Set the output voltage range.
FREQuency:
Set the output frequency value in Hz.
CURRent
This command sets the rms current limit of the output. If the output current exceeds this limit,
the output latches into a disabled state when current limiting occurs. The output voltage
programming is reset to zero when this condition occurs and the output relay state is set to
OFF.
Note that the CURRent command is coupled with the VRANGe. This means that the
maximum current limit that can be programmed at a given time depends on the voltage
range setting in which the unit is presently operating.
Command Syntax
[SOURce:]CURRent <NRf>
Parameters
0 to a value specified by the LIM:CURR
Unit
A (rms amperes)
*RST
Defined by the PON
Examples
CURR 5
Query Syntax
CURRent?
Returned Parameters
<NR2>
Related Commands
VOLT:RANG
FREQuency
This command sets the frequency of the output sinewave.
Command Syntax
[SOURce:]FREQuency <NRf+>
Parameters
range specified by the LIM:FREQ:LOW and LIM:FREQ:HIGH
commands
Unit
Hz (Hertz)
*RST Value
defined by PON
Examples
FREQ 50
Query Syntax
FREQuency?
Returned Parameters
<NR2>
Related Commands
LIM:FREQ;LOW? LIM:FREQ:HIGH?