Overview
|
1-3
Moisture Sensor
Quartz Crystal Microbalance (QCM)
The 3050-RM analyzer uses a quartz crystal microbalance sensor to mea-
sure water vapor concentration. This is a piezoelectric device that is capa-
ble of extremely sensitive mass measurement. To make the QCM sensitive
to water vapor the surface is coated with a thin film of hygroscopic mate-
rial. When gas flows through the sensor, the coating absorbs or desorbs
moisture, changing the natural resonance frequency of the crystal.
Gas measurement cycle
The gas measurement cycle is divided into two periods.
During the first period of the cycle, dry reference gas flows through
the sensor assembly and the frequency is recorded.
During the second period of the cycle, sample gas flows through the
sensor and a second measurement of the sensor frequency is recorded.
The difference between these two frequency measurements is known
as the delta frequency (
∆
ƒ). The
∆
ƒ value is a function of the moisture
concentration in the sample gas.
Gas Flow
Sample system
The 3050-RM Moisture Analyzer sample system enables the analyzer to make
reliable water vapor measurements at concentrations down to 0.1 PPMV.
The sample system serves several purposes.
1. It controls the sample conditions at the sensor assembly (e.g., flow
rate, temperature, pressure). Hence, the QCM-based sensor is isolated
from variations in the ambient conditions that would otherwise alter
the response characteristics of the sensor.
2. The sample system is designed to continuously switch the sensor as-
sembly between the sample gas and a dry reference gas. This sample
switching scheme yields both higher sensitivity and reduced drift in
the sensor readings.
Summary of Contents for 3050-RM
Page 11: ... xi ...
Page 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Page 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Page 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Page 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Page 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Page 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Page 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...