User Interface
|
4-5
Range Key
Use the Range Key to define your 4 to 20 mA output setting.
Use this setting to scale the 4 to 20 mA output. The 4 to 20 mA output is
proportional to the moisture readings.
Maximum Conc.
Enter the moisture reading equivalent to 20 mA.
Minimum Conc.
Enter the moisture reading equivalent to 4 mA.
Example
:
To scale the analog output to represent a 0 to 100 PPMV range enter the
following settings:
Maximum Conc.
100
Minimum Conc.
0
RANGE
Maximum Conc
Minimum Conc
Enter
Value
Enter
Value
Figure 4-5.
Main menu for Range key.