4-8 | 3050-RM Moisture Analyzer
Sample Gas
This list allows you to specify the sample gas you want to analyze.
Select Gas
Press Enter to move to the listing of sample gases. Scroll up/down the list
to select the correct sample gas.
Custom gas
If you press the “Custom Gas ” selection, you will be prompted to enter
a Gas K Factor. The Gas K Factor is a parameter used by the mass flow
meter in the Model 3050-RM to adjust for the molar heat capacities of dif-
ferent gases. Nitrogen is used as the reference gas for the K factors, and
its value is defined as 1.000. To estimate the K factor for another gas, take
the ratio of the molar heat capacity (constant pressure) of nitrogen to that
of the sample gas. For example, the K factor of argon would be estimated
as follows:
Molar heat capacity of nitrogen
= 29.12 JK
-1
mole
-1
Molar heat capacity of argon
= 20.79 JK
-1
mole
-1
K factor
= 29.12 JK
-1
mole
-1
/20.79 JK
-1
mole
-1
K factor
= 1.401
Config
Sample Gas
Select Gas
Air
Argon
Butane
CarbonDioxide
Custom Gas
Ethane
Ethene
Freon12
Freon22
Freon32
Freon114
Freon116
Freon125
Freon134A
Freon152
Helium
Hydrogen
Krypton
NaturalGas
Neon
Nitrogen
Oxygen
Propane
Propene
SF6
Xenon
Gas K Factor
Sensor Saver
Enable
Disable
Gas Saver
Enable
Disable
Summary of Contents for 3050-RM
Page 11: ... xi ...
Page 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Page 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Page 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Page 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Page 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Page 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Page 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...