Serial Communications
|
5-1
SERIAL COMMUNICATIONS
Protocol
The protocol used for the 3050-RM Moisture Analyzer is a master/slave,
command/response communication with the master initiating all trans-
fers. The message format for transfers initiated by the master is seen in
Figure 5-1.
• All elements of this protocol are sent as ASCII characters.
• The Start Flag is the character ‘>’, $3E.
• The End Flag is a carriage return, $0D.
• The Node address is a two byte hexadecimal ASCII value in the range
$00 through $EF.
• Addresses $F0 through $FF are reserved.
• The command byte is a single ASCII character.
• The Data field is of variable length and contains ASCII characters.
• The checksum is a two byte ASCII hexadecimal field representing the
modulo 256 sum of Node Address, Command Byte, and Data fields
(excludes the start flag).
• A master can issue a wildcard in lieu of a checksum.
Figure 5-1.
Message format for transfers
initiated by the master.
Summary of Contents for 3050-RM
Page 11: ... xi ...
Page 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Page 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Page 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Page 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Page 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Page 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Page 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...