6-6 | 3050-RM Moisture Analyzer
Setup Tab
Use the Setup tab to define analyzer parameters.
Figure 6-4.
Setup tab.
Gas
Select the gas being sampled.
Dewpoint Temp - If you select dewpoint temp, the Process Pressure box
is activated and you must enter a process pressure parameter in kPa.
Units
Select the unit of measurement.
Sensor Saver check box
Check to enable Sensor Saver. This feature allows the analyzer to operate
with a slow cycle time to maximize cell life.
Gas Saver check box
Check to enable Gas Saver. This feature allows you to minimize gas usage.
4-20 mA Output
Set up your analog output range.
20 mA
Enter your high analog output limit.
4 mA
Enter your low analog output limit.
Hold during Verify Check this box to hold analyzer output during verifica-
tion.
Alarm Output
Set up the limits for the concentration alarm.
Enable
Check this box to enable the concentration alarm.
High Limit
Enter the high limit for the concentration alarm.
Low Limit
Enter the low limit for the concentration alarm.
To Save settings on the Setup tab, click Apply.
NOTE
Summary of Contents for 3050-RM
Page 11: ... xi ...
Page 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Page 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Page 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Page 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Page 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Page 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Page 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...