Service & Parts
|
8-11
!
CAUTION
Installing the new dryer / trap
Minimize the time that the dryer is exposed to the atmosphere after
removing the protective caps. AMETEK recommends that the dryer
be exposed for less than a minute.
1. Insert the replacement dryer into the dryer port on the analyzer. Use
the new gaskets provided with the dryer in each of the fittings.
2. Tighten the fittings until they are finger tight.
3 Install the dryer mounting screws to secure the dryer to the bracket.
4. Tighten both of the fittings so that the nuts are 1/8
th
of a turn past fin-
ger tight.
5. If you have a second dryer or trap. Repeat Steps 1 through 4 for the
second dryer or trap.
6. Place the dryer compartment cover onto the rear of the analyzer and
secure with the four captured screws.
7. Enter the production code for the new dryer.
The dryer production code is entered only once even if your analyzer
is equipped with two dryers.
8. Power up the analyzer. Wait until the boot-up sequence is complete.
- Press
TEST/CONFIG.
Scroll down to “Production Codes” and press Enter.
- Select “Reference Dryer” and press Enter to enter the production
code. Press Enter.
- Select “Dryer Config” and choose either Single (for one dryer ana-
lyzer) or Double, if your analyzer is equipped with two dryers.
- Press
Cancel several times to reverse through the menu and re-
turn to the normal operating screen.
!
CAUTION
Summary of Contents for 3050-RM
Page 11: ... xi ...
Page 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Page 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Page 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Page 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Page 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Page 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Page 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...