8-12 | 3050 RM Moisture Analyzer
Maintenance of the Electronic Module
The electronics module was designed to easily identify and service the
boards and power supplies used in the Model 3050 RM Analyzer. These
components include:
• MCU Assembly
• Analog Interface Assembly
• Power Supply
• Fan
Removing the Electronic Module from the 3050 RM
Disconnect all power to the analyzer before removing the electronic
module.
Tools required
• 1/4” nut driver
• 1/16” hex wrench
• Phillips and slotted screwdrivers
Procedure for removing the electronics module
Use Figure 8-9 as a reference for locating the items described in this proce-
dure.
• Locate the four (4) Phillips head screws on the electronics module.
• Using the appropriate tool, remove the mounting hardware.
• With all the interconnecting wiring and tubing still attached, remove
the electronic module from the 3050 RM assembly by pulling up and
slightly towards the front of the analyzer.
• Once you remove the electronics module from the assembly, you can
place it on the analyzer cabinet as a resting fixture. However, take
care not to scratch the analyzer cabinet with the electronics module
(a piece of cardboard can be placed under the electronics module to
protect the cabinet).
All of the components are now accessible for maintenance or replacement.
!
CAUTION
Summary of Contents for 3050-RM
Page 11: ... xi ...
Page 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Page 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Page 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Page 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Page 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Page 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Page 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...