2-2
SECTION 2 – OPERATION
Touchscreen Controls
Navigation Keys
Navigation Keys allow the user to change the Content Region to display the different views
such as Home (pictured), Plot, Schedule, etc.
Content Region
This area displays the appropriate content based on which Navigation Key is active.
Instrument Control “Soft” Keys
This area displays instrument-specific “soft” keys. The button will glow red when the key is
active (as displayed above, the Home screen is active).
Fill:
When active, oil is pushed from the Reservoir to the Chamber. This button must be
active to pressurize the Chamber.
Drain:
When active, oil is pushed from the Chamber back to the Reservoir. This button must
be inactive to pressurize the Chamber.
Pump:
This button will activate/deactivate the pump used to pressurize the Chamber.
Pressing this button when the Pressure Controller is active has no effect.
Press Release:
This button will open the Chamber to the Reservoir through a depressurizing
capillary tube. Pressing this button when the Pressure Controller is active has no effect.
Navigation Keys
Content
Region
Instrument Control “Soft” Keys
Summary of Contents for Chandler Engineering 1910
Page 4: ...T 2 TABLE OF CONTENTS This page is intentionally left blank ...
Page 10: ...P 6 PREFACE This page is intentionally left blank ...
Page 12: ...1 2 SECTION 1 INSTALLATION This page is intentionally left blank ...
Page 21: ...SECTION 2 OPERATION 2 9 Help Screen The Help Screen displays this manual ...
Page 24: ...2 12 SECTION 2 OPERATION This page is intentionally left blank ...
Page 27: ......
Page 29: ...4 2 SECTION 4 TROUBLESHOOTING GUIDE This page is intentionally left blank ...
Page 31: ...5 2 SECTION 5 REPLACEMENT PARTS This page is intentionally left blank ...
Page 33: ......
Page 38: ...CHANDLER ENGINEERING ...
Page 40: ...P1 P7 P11 PLC1 P3 P18 P12 P5 P6 CHANDLER ENGINEERING ...
Page 41: ...CHANDLER ENGINEERING ...