ReFlex Power™
AC Power Supplies
M380056-01 Rev L
4-73
4.6.20
H
ALL
C
ALIBRATION FOR
V
OLTAGE
R
ANGE
L
OW
(5
KH
Z
U
NIT
)
The following procedure calibrates the hall gain sensor of the power supply for
the
5KHz AC unit only
. This is seven-point calibration procedure.
1. Close the switch between the power supply and the resistive load.
2. Connect the meter between the output and adjustable resistive load.
3. Set the voltmeter to read the AC current.
4. Ensure that a resistive load is connected to the output.
5. Issue clear and reset commands:
*CLS
*RST<n>
6. Set the voltage range of the power supply to low:
SOUR<n>:VOLT:RANGE 0
7. Turn the unit on:
OUTP<n>:STATE 1
8. Set the hall calibration data to the default value:
CAL<n>:OUTP:FREQ:CURR:DEFAULT
9. Set the output voltage to 120 Volts:
SOUR<n>:VOLT 120
10. Set the output current to 6.3 Amps:
SOUR<n>:CURR 6.3
11. Turn on the resistive load and adjust the load until the output voltage is 110 Volts,
this will ensure that the AC supply is in Constant Current Mode.
12. Set the output frequency to 60Hz
SOUR<n>:FREQ 60
13. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL<n>:OUTP:FREQ:CURR1 <value-1>
14. Set the output frequency to 500Hz
SOUR<n>:FREQ 500
Summary of Contents for Elgar ReFlex Power
Page 1: ...M380056 01 Rev L www programmablepower com ReFlex Power Operation Manual...
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank...
Page 8: ...iv This page intentionally left blank...
Page 10: ...vi This page intentionally left blank...
Page 21: ...M380056 01 Rev L xvii This page intentionally left blank...
Page 22: ......
Page 85: ...ReFlex Power Controller ReFlex Power 3 22 M380056 01 Rev L This page intentionally left blank...
Page 89: ...AC Power Supplies ReFlex Power 4 4 M380056 01 Rev L Figure 4 1 ACPS Module Front Panel 875VA...
Page 120: ...ReFlex Power AC Power Supplies M380056 01 Rev L 4 35 Figure 4 11 Parallel Output Configuration...
Page 147: ...AC Power Supplies ReFlex Power 4 62 M380056 01 Rev L RST n...
Page 157: ...AC Power Supplies ReFlex Power 4 72 M380056 01 Rev L 26 Reset the unit RST n...
Page 274: ...ReFlex Power DC Power Supplies M380056 01 Rev L 6 71 18 Turn off the output RST n...
Page 275: ...DC Power Supplies ReFlex Power 6 72 M380056 01 Rev L This page intentionally left blank...
Page 279: ...Active Loads ReFlex Power 7 4 M380056 01 Rev L Figure 7 1 LPAL 375W Front Panel...
Page 280: ...ReFlex Power Active Loads M380056 01 Rev L 7 5 Figure 7 2 HPAL 750W Front Panel...
Page 281: ...Active Loads ReFlex Power 7 6 M380056 01 Rev L Figure 7 3 Typical Active Load Rear Panel...
Page 359: ...Active Loads ReFlex Power 7 84 M380056 01 Rev L This page intentionally left blank...