ReFlex Power™
Fixed Power DC Power Supply
M380056-01 Rev L
5-5
5.3
SPECIFICATIONS
All specifications are valid after a 30-minute warm-up time. Module can be used
at reduced accuracy immediately following power-on. All accuracy specifications
are plus/minus variations around the nominal parameter values, unless otherwise
noted. Unless otherwise specified, requirements apply at the front panel output
terminals, and with remote sense leads connected to front panel terminals.
Performance specifications are valid following calibration at 25 °C, +/- 5 °C
5.3.1
P
RODUCT
M
ATRIX
Model
Output
Current Rating
Power Rating
FPDC
+5V DC
10 A
50W
-5V DC
5 A
25W
+15V DC
3 A
45W
-15V DC
3 A
45W
5.3.2
O
UTPUT
C
HARACTERISTICS
V
OLTAGE
A
CCURACY
+5VDC and –5 VDC: +/- 52 mV
+15VDC and –15VDC: +/-200 mV
All at 25 °C, +/-5 °C
V
OLTAGE
R
IPPLE
/N
OISE
(measured with 20Hz to 20MHz bandwidth)
+5VDC: 150 mV peak-peak
-5VDC: 80 mV peak-peak
+15VDC and –15VDC: 150 mV peak-peak
V
OLTAGE
T
EMPERATURE
D
RIFT
0.02% of full-scale/°C, maximum
V
OLTAGE
L
INE
P
LUS
L
OAD
R
EGULATION
+5VDC and –5VDC: +20 mV, -100 mV
+15VDC and –15VDC: +/- 150 mV
V
OLTAGE
L
OAD
T
RANSIENT
Less than 4% deviation for a 25-75% of full-scale load step and back within DC
regulation limits after 400 usec
Summary of Contents for Elgar ReFlex Power
Page 1: ...M380056 01 Rev L www programmablepower com ReFlex Power Operation Manual...
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank...
Page 8: ...iv This page intentionally left blank...
Page 10: ...vi This page intentionally left blank...
Page 21: ...M380056 01 Rev L xvii This page intentionally left blank...
Page 22: ......
Page 85: ...ReFlex Power Controller ReFlex Power 3 22 M380056 01 Rev L This page intentionally left blank...
Page 89: ...AC Power Supplies ReFlex Power 4 4 M380056 01 Rev L Figure 4 1 ACPS Module Front Panel 875VA...
Page 120: ...ReFlex Power AC Power Supplies M380056 01 Rev L 4 35 Figure 4 11 Parallel Output Configuration...
Page 147: ...AC Power Supplies ReFlex Power 4 62 M380056 01 Rev L RST n...
Page 157: ...AC Power Supplies ReFlex Power 4 72 M380056 01 Rev L 26 Reset the unit RST n...
Page 274: ...ReFlex Power DC Power Supplies M380056 01 Rev L 6 71 18 Turn off the output RST n...
Page 275: ...DC Power Supplies ReFlex Power 6 72 M380056 01 Rev L This page intentionally left blank...
Page 279: ...Active Loads ReFlex Power 7 4 M380056 01 Rev L Figure 7 1 LPAL 375W Front Panel...
Page 280: ...ReFlex Power Active Loads M380056 01 Rev L 7 5 Figure 7 2 HPAL 750W Front Panel...
Page 281: ...Active Loads ReFlex Power 7 6 M380056 01 Rev L Figure 7 3 Typical Active Load Rear Panel...
Page 359: ...Active Loads ReFlex Power 7 84 M380056 01 Rev L This page intentionally left blank...