ReFlex Power™
DC Power Supplies
M380056-01 Rev L
6-65
MEAS[2000+n11]:VOLT?
MEAS[2000+n11]:CURR?
13. Ensure that the readback values of the output parameters are
within specifications of the actual values.
14. To turn off the output and break down the group with the following
command:
*RST[n]
6.6
CALIBRATION
6.6.1
S
COPE
Procedures are provided in the following sections for calibration of the ReFlex
Power™ DC power supply modules: RFP-D1016-021-XXXX, RFP-D1065-5A1-
XXXX, RFP-D2033-030-XXXX, RFP-D2050-020-XXXX, RFP-D2050-025-XXXX,
RFP-D2120-8A3-XXXX, and RFP-D2450-2A3-XXXX
.
Calibration is easily performed
through the Controller module interface with SCPI commands, requiring only a
voltmeter and current shunt.
WARNING
The calibration procedures are performed with the output of the DC power supply
module energized. Do not touch any of the output connections which could be at
hazardous potentials. Calibration must be performed by qualified personnel who
understand the associated hazards and could appropriately contend with them.
CAUTION
If calibration is not performed properly, functional problems could arise, preventing
the DC power supply module from operating properly.
Summary of Contents for Elgar ReFlex Power
Page 1: ...M380056 01 Rev L www programmablepower com ReFlex Power Operation Manual...
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank...
Page 8: ...iv This page intentionally left blank...
Page 10: ...vi This page intentionally left blank...
Page 21: ...M380056 01 Rev L xvii This page intentionally left blank...
Page 22: ......
Page 85: ...ReFlex Power Controller ReFlex Power 3 22 M380056 01 Rev L This page intentionally left blank...
Page 89: ...AC Power Supplies ReFlex Power 4 4 M380056 01 Rev L Figure 4 1 ACPS Module Front Panel 875VA...
Page 120: ...ReFlex Power AC Power Supplies M380056 01 Rev L 4 35 Figure 4 11 Parallel Output Configuration...
Page 147: ...AC Power Supplies ReFlex Power 4 62 M380056 01 Rev L RST n...
Page 157: ...AC Power Supplies ReFlex Power 4 72 M380056 01 Rev L 26 Reset the unit RST n...
Page 274: ...ReFlex Power DC Power Supplies M380056 01 Rev L 6 71 18 Turn off the output RST n...
Page 275: ...DC Power Supplies ReFlex Power 6 72 M380056 01 Rev L This page intentionally left blank...
Page 279: ...Active Loads ReFlex Power 7 4 M380056 01 Rev L Figure 7 1 LPAL 375W Front Panel...
Page 280: ...ReFlex Power Active Loads M380056 01 Rev L 7 5 Figure 7 2 HPAL 750W Front Panel...
Page 281: ...Active Loads ReFlex Power 7 6 M380056 01 Rev L Figure 7 3 Typical Active Load Rear Panel...
Page 359: ...Active Loads ReFlex Power 7 84 M380056 01 Rev L This page intentionally left blank...