DC Power Supplies
ReFlex Power™
6-68
M380056-01 Rev L
CAL[n]:OUTP:VOLT:COUNTS 14202
9. Measure the output voltage with the external voltmeter. Enter this as value-2 with
the following command:
CAL[n]:OUTP:VOLT:FIVEPOINT 2 <value->2
10. Program the output voltage to the third value near 50% of full-scale,
count = 48852*0.50+768 = 25194):
CAL[n]:OUTP:VOLT:COUNTS 25194
11. Measure the output voltage with the external voltmeter. Enter this as value-3 with
the following command:
CAL[n]:OUTP:VOLT:FIVEPOINT 3 <value-3>
12. Program the output voltage to the fourth value near 72.5% of full-scale,
(count = 48852*0.725+768 = 36185):
CAL[n]:OUTP:VOLT:COUNTS 36185
13. Measure the output voltage with the external voltmeter. Enter this as value-4 with
the following command:
CAL[n]:OUTP:VOLT:FIVEPOINT 4 <value-4>
14. Program the output voltage to the fifth value near 95% of full-scale,
(count = 48852*0.95+768 = 47177):
CAL[n]:OUTP:VOLT:COUNTS 47177
15. Measure the output voltage with the external voltmeter. Enter this as value-5 with
the following command:
CAL[n]:OUTP:VOLT:FIVEPOINT 5 <value-5>
16. Save the calibration results with the following commands:
CAL[n]:UNLOCK “6867”
CAL[n]:STORE
CAL[n]:LOCK
(The entered voltage data could be verified with the following query:
CAL[n]:OUTP:VOLT:FIVEPOINT?
)
17. Turn off the output:
Summary of Contents for Elgar ReFlex Power
Page 1: ...M380056 01 Rev L www programmablepower com ReFlex Power Operation Manual...
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank...
Page 8: ...iv This page intentionally left blank...
Page 10: ...vi This page intentionally left blank...
Page 21: ...M380056 01 Rev L xvii This page intentionally left blank...
Page 22: ......
Page 85: ...ReFlex Power Controller ReFlex Power 3 22 M380056 01 Rev L This page intentionally left blank...
Page 89: ...AC Power Supplies ReFlex Power 4 4 M380056 01 Rev L Figure 4 1 ACPS Module Front Panel 875VA...
Page 120: ...ReFlex Power AC Power Supplies M380056 01 Rev L 4 35 Figure 4 11 Parallel Output Configuration...
Page 147: ...AC Power Supplies ReFlex Power 4 62 M380056 01 Rev L RST n...
Page 157: ...AC Power Supplies ReFlex Power 4 72 M380056 01 Rev L 26 Reset the unit RST n...
Page 274: ...ReFlex Power DC Power Supplies M380056 01 Rev L 6 71 18 Turn off the output RST n...
Page 275: ...DC Power Supplies ReFlex Power 6 72 M380056 01 Rev L This page intentionally left blank...
Page 279: ...Active Loads ReFlex Power 7 4 M380056 01 Rev L Figure 7 1 LPAL 375W Front Panel...
Page 280: ...ReFlex Power Active Loads M380056 01 Rev L 7 5 Figure 7 2 HPAL 750W Front Panel...
Page 281: ...Active Loads ReFlex Power 7 6 M380056 01 Rev L Figure 7 3 Typical Active Load Rear Panel...
Page 359: ...Active Loads ReFlex Power 7 84 M380056 01 Rev L This page intentionally left blank...