18
2017-08-11
125614
04
4.2.1
Connecting RTDs
See section 10.0; Technical specifications for supported types of
RTDs. The calibrator accepts 2-, 3-, and 4-wire inputs, with the 4-wire
input being the most accurate.
4.2.2
Connecting Thermocouples and Voltages Sources
See section 10.0; Technical specifications for supported types of
thermocouples. The multi-scanner has a cold junction compensation
function. Normally this function should be ON and the actual
temperature of the thermocouple can be measured.
With CJC OFF, the multi-scanner will measure the difference
between the thermocouple at the junction and at its TC input terminal.
2-wire
3-wire
4-wire
A -
B -
C -
D -