30
2017-08-11
125614
04
6.3.17 TEMP_UNIT
TEMP_UNIT <temp. unit>
<temp. unit> = CEL, FAR, KEL
Sets temperature unit in all connected units. Default setting at power
up is CEL.
6.3.18 TEMP_UNIT?
Returns temperature unit: <temp. unit> = CEL, FAR, KEL
If the connected units have different configurations “ERROR” will be
returned.
6.3.19 CH_CONF
Configures the channels.
CH_CONF DCV (@<ch_list>)
Sets the selected for measuring voltage.
Example: CH_CONF DCV (@2)
CH_CONF DCV, <Range low> V, <Range high> V, <Temperature
low> <temp. unit>, <Temperature high> <temp. unit>
(@<ch_list>)
Sets the selected for measuring voltage transmitter (T/V)
<Range low> = V_LOW_XX = 0.000 [V] (default value)
<Range high> = V_HIGH_XX = 10.000 [V] (default value)
<Temperature low> = V_TEMP_LOW_XX = 0 [°C] (default value)
<Temperature high> = V_TEMP_HIGH_XX = 100 [°C] (default value)
CH_CONF MA (@<ch_list>)
Sets the selected channels for measuring current
CH_CONF MA, <Range low> A, <Range high> A, <Temperature
low> <temp. unit> , <Temperature high> <temp. unit>
(@<ch_list>)
Sets the selected channels for measuring current transmitter (T/I)